The aim of this study was to investigate the relationship between partial electrical reset (PER) and CT scan parameters (tube voltage, tube current, rotation time, and product of tube current and rotation time in mAs). A cardiac resynchronization therapy pacemaker (Insync 8040, Medtronic Inc., Tokyo) and 320 area detector CT scanner (Aquilion ONE, Toshiba medical systems, Otawara, Japan) with volume scan were used. The pacemaker was put in DDD mode. The PERs were interpreted using both the programmer's wave forms and error messages. The exposure was repeated 5 times per CT setting. The pacemaker was placed on the anterior wall and upper side of a chest phantom. Each CT scan was performed using the following parameters: tube voltage of 80, 100, 120, and 135 kV; tube current of 50-550 mA; and rotation time of 0.35-1.5 s. PERs were observed at 100, 120, and 135 kV, and more PERs were observed as the tube voltage increased. The PER tube current decreased as the rotation time was increased. In contrast, the PER tube current and rotation time product (mAs) increased as the rotation time was increased. More specifically, the radiation dose rate was the affected factor of the PERs. To avoid PER of pacemakers, CT scan parameters with lower radiation dose rates (low rather than high tube current and rotational time) is recommended. In conclusion, our results will help with CT scans of patients who have implantable cardiac devices (included pacemakers and cardioverter defibrillators).

Download full-text PDF

Source
http://dx.doi.org/10.6009/jjrt.67.648DOI Listing

Publication Analysis

Top Keywords

tube current
28
rotation time
24
tube voltage
16
current rotation
16
tube
11
electrical reset
8
implantable cardiac
8
cardiac devices
8
voltage tube
8
scan parameters
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!