Background And Aim Of The Study: The study aim was to evaluate the hemodynamic performance of the St. Jude Medical Epic Supra bioprosthesis during the early six-month follow up period, and to confirm the safety and efficacy of the valve by collecting details of adverse events and NYHA functional classification.

Methods: Fifty-seven patients undergoing aortic valve replacement (AVR) with the Epic Supra valve between September 2007 and January 2009 at three centers in Canada were evaluated for the study. The subjects were monitored preoperatively, at discharge, and at six months postoperatively. Echocardiographic data were available from 50 subjects at the six-month follow up. In order to prevent observer variability, all echocardiograms were sent to an independent Echocardiography Core Laboratory (ECL) for interpretation of the data.

Results: The mean subject age was 74 years. Concomitant coronary artery bypass grafting (CABG) was performed in 44% of the procedures. The mean pressure gradients were 11.2, 12.5, 10.8, 8.4 and 11.3 mmHg, respectively, for valves sized 19 mm (n = 2), 21 mm (n = 20), 23 mm (n = 22), 25 mm (n = 5) and 27 mm (n = 1). The average effective orifice areas (EOAs) were 1.44, 1.57, 1.69, 1.93 and 1.81 cm2 for the .valves sized 19, 21, 23, 25 and 27 mm, respectively.

Conclusion: The results of the six-month echocardiographic follow up indicated that the Epic Supra valve offered excellent hemodynamic performance in the 21, 23 and 25 mm sizes. However, additional data are still required for the 19 and 27 mm valves to characterize their performance. The mean gradients and EOA-values were comparable to those of other supra-annular stented tissue valves. The EOA index indicated an absence of prosthesis-patient mismatch, with values in all subjects at or near 0.85 cm2/m2. The percentage of subjects without aortic insufficiency (AI) at follow up was 92%; only four subjects showed trivial AI.

Download full-text PDF

Source

Publication Analysis

Top Keywords

epic supra
16
hemodynamic performance
12
performance jude
8
jude medical
8
medical epic
8
six-month follow
8
supra valve
8
valves sized
8
valve
5
subjects
5

Similar Publications

(1) Background: This study aimed to analyse the hydrodynamic performance of two generations of self-expanding transcatheter heart valves (THV) as a valve-in-valve (ViV) in different surgical aortic valve (SAV) models under standardised conditions. The nitinol-based Evolut R valve is frequently used in ViV procedures. It is unclear whether its successor, the Evolut PRO, is superior in ViV procedures, particularly considering the previously implanted SAV model.

View Article and Find Full Text PDF

Background And Aims: The durability of surgical aortic valve bioprostheses (SAV) is limited by the calcification of the leaflets, which results in degeneration. In clinical routine, there seems to be substantial variability in the degeneration of specific SAV models. Our study aims to establish an in vitro calcification model for prosthetic valves, characterizing the calcification behavior of different SAVs.

View Article and Find Full Text PDF

The St. Jude Medical Epic Supra valve is a porcine bioprosthesis designed for complete supraannular implantation. No report has shown the hemodynamic performance and clinical outcomes of aortic valve replacement with the Epic Supra valve for severe aortic stenosis in a Japanese cohort.

View Article and Find Full Text PDF
Article Synopsis
  • * Over an average follow-up of 6.2 years, the survival rates were high at 91% after 2 years and 88% after 5 years, with 86% and 83% respectively avoiding major cardiovascular events.
  • * The valve showed excellent durability, with no significant deterioration in function over time and high rates of freedom from severe structural issues, indicating that the Epic Supra valve performs well in the long term.
View Article and Find Full Text PDF

Objective: Bioprosthetic surgical aortic valve replacement remains an important treatment option in the era of transcatheter interventions. Real-world outcomes are not well characterized because of limited prospective follow-up studies. We present the 10-year clinical outcomes of Medicare beneficiaries undergoing surgical aortic valve replacement with a contemporary supra-annular porcine valve.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!