The impact of tensioning device mal-positioning on strand tension during anterior cruciate ligament reconstruction.

J Orthop Surg Res

Comparative Orthopaedic Research Surgical Facility, School of Medicine, Flinders University, South Australia, Australia.

Published: June 2011

Background: In order to confer optimal strength and stiffness to the graft in Anterior Cruciate Ligament (ACL) reconstruction, the maintenance of equal strand tension prior to fixation, is desired; positioning of the tensioning device can significantly affect strand tension This study aimed to determine the effect of tensioning device mal-positioning on individual strand tension in simulated cadaveric ACL reconstructions.

Methods: Twenty cadaveric specimens, comprising bovine tibia and tendon harvested from sheep, were used to simulate ACL reconstruction with a looped four-strand tendon graft. A proprietary tensioning device was used to tension the graft during tibial component fixation with graft tension recorded using load cells. The effects of the tensioning device at extreme angles, and in various locking states, was evaluated.

Results: Strand tension varied significantly when the tensioning device was held at extreme angles (p < 0.001) or in 'locked' configurations of the tensioning device (p < 0.046). Tendon position also produced significant effects (p < 0.016) on the resultant strand tension.

Conclusion: An even distribution of tension among individual graft strands is obtained by maintaining the tensioning device in an unlocked state, aligned with the longitudinal axis of the tibial tunnel. If the maintenance of equal strand tension during tibial fixation of grafts is important, close attention must be paid to positioning of the tensioning device in order to optimize the resultant graft tension and, by implication, the strength and stiffness of the graft and ultimately, surgical outcome.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3133604PMC
http://dx.doi.org/10.1186/1749-799X-6-33DOI Listing

Publication Analysis

Top Keywords

tensioning device
36
strand tension
24
tension
10
device
9
device mal-positioning
8
anterior cruciate
8
cruciate ligament
8
strength stiffness
8
stiffness graft
8
acl reconstruction
8

Similar Publications

Background/aims: Levator ani syndrome (LAS) is a pathology that is both difficult to diagnose and treat. The effectiveness of current treatments, which are mostly administered with special devices in advanced healthcare centers, is controversial. The aim of the study is to investigate the effects of glyceryl trinitrate, which is easily accessible and can be applied by the patient, in the treatment of LAS.

View Article and Find Full Text PDF

Objectives: This study aims to analyze the biomechanics of three kinds of rigid internal fixation methods for condylar head fractures.

Methods: A three dimensional finite element model of the normal mandible was constructed. It was then used to prepare condylar head fracture finite element model and three kinds of rigid internal fixation finite element model (unilateral tension screw, bilateral tension screw, tension screw+titanium plate).

View Article and Find Full Text PDF

For gastric leiomyomas measuring ≥5 cm, endoscopic resection is necessary. The larger size of these tumors significantly impairs the resection field of view, increasing the risk of intraoperative bleeding and perforation and potentially leading to incomplete tumor removal. The combination of dental floss and tissue clip traction techniques is commonly used for resecting mucosal lesions but is rarely reported for submucosal tumors.

View Article and Find Full Text PDF

Objective: The treatment of comminuted inferior pole patellar fractures has long posed a challenge for orthopedic surgeons. This study aims to compare the biomechanical stability and clinical efficacy of Kirschner wire tension band combined with anchor cross-suture fixation versus traditional partial patellectomy in the treatment of comminuted inferior pole patellar fractures.

Methods: A retrospective analysis was conducted on 14 patients who underwent Kirschner wire tension band combined with anchor cross-suture fixation (Group A) in our department of orthopedics from September 2020 to April 2022.

View Article and Find Full Text PDF

Purpose: An assistive technology ecosystem requires co-ordinated collaboration between policy, products, provision processes, and personnel. A needs assessment was conducted of the assistive technology sector in Queensland to examine these components from the perspective of all key stakeholders. This paper reports the findings of the expressed needs of the assistive technology sector as experienced by the AT advisors and suppliers.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!