  • This study introduces a new approach to improve intraarterial thrombolysis using an ionic polymer-metal composite (IPMC) actuator to enhance the dissolution of blood clots.
  • The research involved testing various sizes of IPMC actuators in a vascular phantom model, finding that specific conditions (3 V, 5 Hz, for 5 minutes) significantly increased the effectiveness of thrombolysis compared to standard treatments.
  • The study concludes that the developed vibrating polymer actuator probe shows promise for future applications in intravascular thrombolysis and suggests further testing in animal models.

Article Abstract

In this study, we propose a new method for enhancement of intraarterial thrombolysis using an ionic polymer-metal composite (IPMC) actuator. The purpose of this study was to test the mechanical thrombolysis efficiency of IPMC actuators and evaluate the endovascular vibrating polymer actuator probe for mechanical thrombolysis in a phantom model; 2 × 1 × 15 mm (2 mm in width, 1 mm in thickness, and 15 mm in length) and 0.8 × 0.8 × 10 mm (0.8 mm in width, 0.8 mm in thickness, and 10 mm in length) IPMC actuators were fabricated by stacking five and four Nafion-117 films, respectively. We manufactured the endovascular vibrating polymer actuator probe, for which thrombolysis efficiency was tested in a vascular phantom. The phantom study using 2 × 1 × 15 mm IPMC actuators showed that 5 Hz actuation is the optimal frequency for thrombolysis under both 2 and 3 V, when blood clot was not treated with rtPA, and when exposed to rtPA, IPMC actuators under the optimized condition (3 V, 5 Hz, and 5 min) significantly increased the thrombolysis degree compared with control and other experimental groups (p < 0.05). In addition, 0.8 × 0.8 × 10 mm IPMC actuators also revealed a significantly higher thrombolysis degree under the optimized condition than the control and rtPA only groups (p < 0.05). Finally, the fabricated probe using 0.8 × 0.8 × 10 mm IPMC actuators also incurred higher thrombolysis degree under the optimized condition than the control and rtPA only groups (p < 0.05). A vibrating polymer actuator probe is a feasible device for intravascular thrombolysis, and further study in an animal model is warranted.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MAT.0b013e31822188ceDOI Listing

Publication Analysis

Top Keywords

ipmc actuators
24
vibrating polymer
16
polymer actuator
16
actuator probe
16
endovascular vibrating
12
mechanical thrombolysis
12
optimized condition
12
thrombolysis degree
12
groups 005
12
thrombolysis
10

Similar Publications

Ionic Polymer-Metal Composite (IPMC) actuators have garnered significant scientific attention in robotics and artificial muscles for their ability to operate at low voltage, high strain capacity, and lightweight construction. The lack of uniform bending in IPMC actuators undermines their control precision and restricts their range of potential applications. This study utilized the unique properties of nanoscale materials and Polyvinyl alcohol (PVA) to develop a membrane for soft robotic bending actuation.

View Article and Find Full Text PDF

Ionic polymer metal composite (IPMC) is regarded as the mainstream application material for achieving flexible driving technology in various engineering fields. In this article, aiming at the non-independence of the current IPMC electric driving method, an IPMC optical-controlled flexible driving method based on the photoinduced effects of lanthanum-modified lead zirconate titanate (PLZT) ceramic is proposed. To this end, a mathematical model for IPMC optical controlled flexible driving is built on the basis of the photovoltaic characteristic of PLZT ceramic, and the driving performance is experimentally analyzed through different lengths of IPMC under the excitation of different direct currents and light intensities.

View Article and Find Full Text PDF

IPMC-based actuators: An approach for measuring a linear form of its static equation.

Heliyon

February 2024

Department of Mechanics, Electrical Power and Computer, Science and Research Branch, Islamic Azad University, Tehran, Iran.

Ionic Polymer-Metal Composites (IPMCs) are smart materials used as actuators, sensors, and energy harvesters. They are known as a subset of Electroactive Polymers (EAPs). IPMCs structure is layered with one polymer layer and two of electrodes.

View Article and Find Full Text PDF

Ionic Polymer-Metal Composites: From Material Engineering to Flexible Applications.

Acc Chem Res

January 2024

Institute of Functional Nano & Soft Materials, Soochow University, Suzhou, Jiangsu 215123, China.

ConspectusIonic polymer-metal composites (IPMCs) are one kind of artificial muscles that can realize energy conversions in response to external stimulus with merits of lightweight, scalability, quick response, and flexibility and have been treated as an important platform in artificial intelligence, such as bionic robotics, smart sensors, and micro-electromechanical systems. It is well-known that IPMC devices are mainly composed of one electrolyte layer laminated with symmetric electrode layers and realize energy conversion based on ion migration and redistribution inside the devices. However, several critical issues have greatly impeded the practical applications of IPMC devices, including metal electrode cracks, metal-polymer interface detachment, and water loss in the electrolyte.

View Article and Find Full Text PDF

Gold nanowire mesh electrode for electromechanical device.

Sci Rep

October 2023

Research Center for Macromolecules and Biomaterials, National Institute for Materials Science, 1-1 Namiki, Tsukuba, Ibaraki, 305-0044, Japan.

Ionic polymer-metal composite (IPMC) actuators were prepared with Nafion film as the ionic polymer and gold nanowire (Au-NW) mesh film as the metal electrodes by hot-pressing, which shortened preparation time within 1 h. As a reference, IPMC actuator consisting of Nafion film and gold foil (Au-foil) was also prepared. Au-NW mesh film can be an electrode with thinner (about 150 nm) and lower surface resistivity (about 0.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!