Fast radiofrequency flip angle calibration by Bloch-Siegert shift.

Magn Reson Med

Imaging Technologies Laboratory, General Electric Global Research, Garching b. Munchen, Germany.

Published: November 2011

In a recent work, we presented a novel method for B 1+ field mapping based on the Bloch-Siegert shift. Here, we apply this method to automated fast radiofrequency transmit gain calibration. Two off-resonance radiofrequency pulses were added to a slice-selective spin echo sequence. The off-resonance pulses induce a Bloch-Siegert phase shift in the acquired signal that is proportional to the square of the radiofrequency field magnitude B(1) (2) . The signal is further spatially localized by a readout gradient, and the signal-weighted average B(1) field is calculated. This calibration from starting system transmit gain to average flip angle is used to calculate the transmit gain setting needed to produce a desired imaging sequence flip angle. A robust implementation is demonstrated with a scan time of 3 s. The Bloch-Siegert-based calibration was used to predict the transmit gain for a 90° radiofrequency pulse and gave a flip angle of 88.6 ± 3.42° when tested in vivo in 32 volunteers.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.22902DOI Listing

Publication Analysis

Top Keywords

flip angle
16
transmit gain
16
fast radiofrequency
8
bloch-siegert shift
8
flip
4
radiofrequency flip
4
angle
4
calibration
4
angle calibration
4
calibration bloch-siegert
4

Similar Publications

Purpose: The single reference variable flip angle sequence with a multi-echo stack of stars acquisition (SR-VFA-SoS) simultaneously measures temperature change using proton resonance frequency (PRF) shift and T-based thermometry methods. This work evaluates SR-VFA-SoS thermometry in MR-guided focused ultrasound in an in vivo rabbit model.

Methods: Simultaneous PRF shift thermometry and T-based thermometry were obtained in a New Zealand white rabbit model (n = 7) during MR-guided focused ultrasound surgery using the SR-VFA-SoS sequence at 3 T.

View Article and Find Full Text PDF

α-Tocopherol (αtoc, vitamin E) is an essential nutrient sufficiently acquired through a balanced diet. This fat-soluble vitamin is most known for its antioxidative properties, however, its fundamental mechanism of action in cellular membranes remains unknown. To this end, we use time-resolved small angle neutron scattering (TR-SANS) and a contrast matching scheme to determine intervesicular exchange (k) and intrabilayer flip-flop (k) rates of αtoc in 1,2-dipalmitoyl-sn-glycero-3-phosphocholine (DPPC) vesicles.

View Article and Find Full Text PDF

Purpose: To provide a fast quantitative imaging approach for a 0.55T scanner, where signal-to-noise ratio is limited by the field strength and k-space sampling speed is limited by a lower specification gradient system.

Methods: We adapted the three-dimensional spiral projection imaging MR fingerprinting approach to 0.

View Article and Find Full Text PDF

Background And Purpose: Magnetic Resonance Imaging is widely used to assess disease burden in multiple sclerosis (MS). This study aimed to evaluate the effectiveness of a commercially available k-nearest neighbors (k-NN) software in quantifying white matter lesion (WML) burden in MS. We compared the software's WML quantification to expert radiologists' assessments.

View Article and Find Full Text PDF

Purpose: Compressed Sensing (CS) is an emerging technique to accelerate MRI acquisitions. The aim of this study was to assess the reliability and accuracy of cartilage thickness measurements in the knee using a CS-enabled isotropic 3D Fast Spin-Echo (FSE) sequence on a 3-T MRI scanner.

Methods: Twenty-eight tibial condyle sections were collected from 14 adult patients who underwent total knee arthroplasty.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!