Selecting the instruments to determine the operational quantities in the neutron fields produced by particle accelerators involves a combination of aspects, which is peculiar to these environments: the energy distribution of the neutron field, the continuous or pulsed time structure of the beam, the presence of other radiations to which the neutron instruments could have significant response and the large variability in the dose rate, which can be observed when moving from areas near the beam line to free-access areas. The use of spectrometric techniques in support of traditional instruments is highly recommended to improve the accuracy of dosimetric evaluations. The multi-sphere or Bonner Sphere Spectrometer (BSS) is certainly the most used device, due to characteristics such as the wide energy range, large variety of active and passive detectors suited for different workplaces, good photon discrimination and the simple signal management. Disadvantages are the poor energy resolution, weight and need to sequentially irradiate the spheres, leading to usually long measurement sessions. Moreover, complex unfolding analyses are needed to obtain the neutron spectra. This work is an overview of the BSS for area monitoring in particle accelerators.