Purpose: To determine the difference in radiation dose between non-enhanced (NECT) and contrast-enhanced (CECT) chest CT examinations contributed by contrast material with different scanner generations with automatic exposure control (AEC).

Methods & Materials: Each 42 adult patients received a NECT and CECT of the chest in one session on a 16-, 64- or 128-slice CT scanner with the same scan protocol settings. However, AEC technology (Care Dose 4D, Siemens) underwent upgrades in each of the three scanner generations. DLP, CTDIvol and image noise were compared.

Results: Although absolute differences in image noise were very small and ranged between 10 and 13 HU for NECT and CECT in median, the differences in image noise and dose (DLP: 16-slice:+2.8%; 64-slice:+3.9%; 128-slice:+5.6%) between NECT and CECT were statistically significant in all groups. Image noise and dose parameters were significantly lower in the most recent 128-slice CT generation for both NECT and CECT (DLP: 16-slice:+35.5-39.2%; 64-slice:+6.8-8.5%).

Conclusion: The presence of contrast material lead to an increase in dose for chest examinations in three CT generations with AEC. Although image noise values were significantly higher for CECT, the absolute differences were in a range of 3 HU. This can be regarded as negligible, thus indicating that AEC is able to fulfill its purpose of maintaining image quality. However, technological developments lead to a significant reduction of dose and image noise with the latest CT generation.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejrad.2011.05.012DOI Listing

Publication Analysis

Top Keywords

image noise
28
nect cect
16
contrast material
12
image
8
radiation dose
8
automatic exposure
8
exposure control
8
16- 64-
8
64- 128-slice
8
cect chest
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!