Purpose: Intensity-modulated radiation therapy (IMRT) may be useful to reduce toxicity in gynecologic cancer patients requiring postoperative pelvic irradiation. This study was undertaken to quantify vaginal wall organ motion during the course of postoperative pelvic irradiation using pelvic IMRT.
Methods And Materials: Twenty-two consecutive patients treated with postoperative pelvic IMRT on helical tomotherapy had fiducial markers placed at the vaginal apex prior to simulation then daily megavoltage computed tomography (CT) scans for positioning. The daily positions of the fiducials were registered and measured in reference to the initial CT scan to quantify the degree of vaginal wall organ motion during the entire course of therapy.
Results: The total motion of the fiducials center of mass (COM) was a median of 5.8 mm (range, 0.6-20.2 mm), and 95% of all COM positions fell within 15.7 mm of their original position. Directional margins of 3.1 mm along the right-left axis, 9.5 mm along the superoinferior axis, and of 12.1 mm along the anteroposterior axis encompassed the vaginal fiducials in 95% of treatments. Mean organ deformation for all patients was 3.9 mm, (range, 0-27.5 mm; standard deviation, 3.1 mm), with significant distortions of greater than 10 mm in 17% of secondary image sets.
Conclusions: These data suggest a planning target volume margin of 16 mm will account for maximal organ motion in the majority of gynecologic patients undergoing postoperative pelvic IMRT, and it may be possible to incorporate directional motion into the planning target volume margin.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ijrobp.2011.01.054 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!