Hydroxylapatite ossicular replacement prostheses: preliminary results.

Laryngoscope

Department of Otolaryngology, School of Medicine, Wright State University, Dayton, Ohio.

Published: July 1990

Hearing results and extrusion rate for 89 consecutive patients receiving hydroxylapatite prostheses for hearing reconstruction were evaluated and compared to a control group of 75 patients who had received homograft bone or Plasti-Pore prostheses. At the 3-month follow-up, there were no cases of extrusion of the hydroxylapatite prostheses. Hearing success was defined as a postoperative air-bone gap of less than or equal to 15 dB for incus prostheses and partial ossicular replacement prostheses or less than or equal to 25 dB for incus-stapes prostheses and total ossicular replacement prostheses. There was a statistically significant difference in the distribution-of-success rate by prosthesis type for the hydroxylapatite. Incus, incus-stapes, and total ossicular replacement prosthesis results were similar, but partial ossicular replacement prosthesis results were poorer. The overall success rate in the hydroxylapatite group was 51.2%; in the control group, it was 60%. Hearing results tended to be better with the hydroxylapatite incus prosthesis than with the homograft bone incus, but poorer with the hydroxylapatite partial ossicular replacement prosthesis than with the Plasti-Pore. Several technical modifications of the prosthesis that may improve results with the hydroxylapatite partial ossicular replacement prosthesis are suggested. Hydroxylapatite middle ear prostheses appear to be a reasonable alternative to other more commonly used prostheses, which carry disadvantages related to the use of human tissue.

Download full-text PDF

Source
http://dx.doi.org/10.1288/00005537-199007000-00004DOI Listing

Publication Analysis

Top Keywords

ossicular replacement
28
partial ossicular
16
replacement prosthesis
16
replacement prostheses
12
prostheses
10
hydroxylapatite
9
hydroxylapatite prostheses
8
prostheses hearing
8
control group
8
homograft bone
8

Similar Publications

Exploring diverse biomaterials and implants in the ear, nose, and throat by understanding adverse effects and post-usage events. Literature was obtained from Scopus, PubMed, Google Scholar, and Web of Science. A comprehensive analysis was conducted on original research studies, case reports, and case series spanning from December 2010 to May 2022.

View Article and Find Full Text PDF

In this study, a method for determining the optimal location and orientation of an implantable piezoelectric accelerometer on the short process of the incus is presented. The accelerometer is intended to be used as a replacement for an external microphone to enable totally implantable auditory prostheses. The optimal orientation of the sensor and the best attachment point are determined based on two criteria-maximum pressure sensitivity sum and minimum loudness level sum.

View Article and Find Full Text PDF

Audiometric results of tympanoplasty with stapedectomy and synthetic total ossicular replacement prosthesis: A STROBE analysis.

Eur Ann Otorhinolaryngol Head Neck Dis

December 2024

Service d'ORL et de chirurgie cervico-faciale, centre hospitalo-universitaire, Dijon, France; Laboratoire ICMUB, UMR CNRS 6302, université de Bourgogne, Dijon, France.

Aim: To evaluate the audiometric results of tympanoplasty with stapedectomy and insertion of a synthetic total ossicular replacement prosthesis (TORP).

Material And Methods: Retrospective observational study conducted on a cohort of 15 patients (16 ears) aged 10 to 58 years (mean: 36 years) with chronic otitis media (COM) and tympanosclerosis (n=9), stapediovestibular ankylosis (n=3), minor aplasia (n=3) or post-traumatic ossicular dislocation (n=1). Treatment consisted in tympanoplasty with stapedectomy and TORP placement by the same operator, between December 1, 2012 and January 30, 2023, in a French university hospital department, with follow-up ranging from 2 to 92 months (mean: 24 months).

View Article and Find Full Text PDF

Investigation of the coupling quality of partial prostheses at the stapes head.

Eur Arch Otorhinolaryngol

November 2024

Carl Gustav Carus Faculty of Medicine, Department of Otorhinolaryngology, Head and Neck Surgery, Ear Research Center Dresden, Technische Universität Dresden, Fetscherstraße 74, 01307, Dresden, Germany.

Article Synopsis
  • The study investigates how the connection quality between different types of prostheses and the stapes head affects the success of partial ossicular chain reconstructions in conductive hearing loss cases.
  • It uses a standardized procedure to compare the coupling quality of various prostheses, including clip and bell types, using 12 temporal bones in a controlled model.
  • Results indicate that all prosthesis types provide effective sound transmission with minimal vibrational loss and distortion, suggesting that while the coupling method may impact practical aspects, it does not significantly affect sound performance.
View Article and Find Full Text PDF

Retrospective analysis of clinical data of 123 patients with atticotomy, exploring the clinical characteristics of patients undergoing atticotomy and the efficacy of hearing reconstruction methods. 123 patients with atticotomy were divided into three groups according to the ossicular chain treatment method: preservation of the ossicular chain group(37 cases), cartilage elevation of stapes group(49 cases), and PORP group(37 cases). The clinical characteristics of patients with atticotomy, preoperative and postoperative hearing levels of the three groups of patients, and postoperative complications were analyzed.

View Article and Find Full Text PDF