Objectives: To evaluate the lowest radiation exposure threshold at which coronary calcium scoring (CCS) remains accurate.

Methods: A prospective study of 43 consecutive eligible patients referred for CCS underwent imaging in accordance with the manufacturer's recommended protocol. Dedicated software was used to generate 8 series of images simulating tube currents ranging from 20 to 300 mA. These images were randomised and read in blinded fashion to determine the lowest tube current at which the CCS remained accurate. The minimum mA was correlated with 6 different patients' biometric parameters: bodyweight, body mass index, AP and lateral thoracic diameters, average thoracic diameter and the scout attenuation coefficient (SAC). The 95% confidence interval for each parameter was used to calculate tube current threshold levels and hence stratified CCS protocols were derived.

Results: Spearman's correlation coefficients of the minimum tube current for the 6 parameters were: 0.66, 0.63, 0.65, 0.74, 0.77 and 0.86 respectively (p < 0.001). SAC offered the largest potential reduction in mean effective dose from 1.86 mSv to 0.88 mSv.

Conclusion: CCS with at least 50% reduction in radiation exposure and below 1 mSv is feasible if CT scout projections are utilised effectively.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00330-011-2159-6DOI Listing

Publication Analysis

Top Keywords

tube current
12
threshold coronary
8
radiation dose
4
dose threshold
4
coronary artery
4
artery calcium
4
calcium score
4
score mdct
4
mdct low
4
low go?
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!