Purpose: In this study, newly formulated XR-RV3 GafChromic film was calibrated with National Institute of Standards and Technology (NIST) traceability for measurement of patient skin dose during fluoroscopically guided interventional procedures.
Methods: The film was calibrated free-in-air to air kerma levels between 15 and 1100 cGy using four moderately filtered x-ray beam qualities (60, 80, 100, and 120 kVp). The calibration films were scanned with a commercial flatbed document scanner. Film reflective density-to-air kerma calibration curves were constructed for each beam quality, with both the orange and white sides facing the x-ray source. A method to correct for nonuniformity in scanner response (up to 25% depending on position) was developed to enable dose measurement with large films. The response of XR-RV3 film under patient backscattering conditions was examined using on-phantom film exposures and Monte Carlo simulations.
Results: The response of XR-RV3 film to a given air kerma depended on kVp and film orientation. For a 200 cGy air kerma exposure with the orange side of the film facing the source, the film response increased by 20% from 60 to 120 kVp. At 500 cGy, the increase was 12%. When 500 cGy exposures were performed with the white side facing the x-ray source, the film response increased by 4.0% (60 kVp) to 9.9% (120 kVp) compared to the orange-facing orientation. On-phantom film measurements and Monte Carlo simulations show that using a NIST-traceable free-in-air calibration curve to determine air kerma in the presence of backscatter results in an error from 2% up to 8% depending on beam quality. The combined uncertainty in the air kerma measurement from the calibration curves and scanner nonuniformity correction was +/- 7.1% (95% C.I.). The film showed notable stability. Calibrations of film and scanner separated by 1 yr differed by 1.0%.
Conclusions: XR-RV3 radiochromic film response to a given air kerma shows dependence on beam quality and film orientation. The presence of backscatter slightly modifies the x-ray energy spectrum; however, the increase in film response can be attributed primarily to the increase in total photon fluence at the sensitive layer. Film calibration curves created under free-in-air conditions may be used to measure dose from fluoroscopic quality x-ray beams, including patient backscatter with an error less than the uncertainty of the calibration in most cases.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3078021 | PMC |
http://dx.doi.org/10.1118/1.3560422 | DOI Listing |
Indian Heart J
January 2025
Department of Cardiology, Sri Jayadeva Institute of Cardiovascular Sciences and Research, Bangalore, India.
Background: Cardiac catheterization via the trans-radial approach (TRA) has shown several advantages over the trans-femoral approach (TFA) but with a concern of higher radiation exposure. Considering the growing experience with TRA, this study compares patient's radiation during coronary angiography using TRA versus TFA.
Methods: This study included consecutive patients undergoing coronary angiogram over a year at tertiary hospital performed by experienced operators through radial or femoral access.
Heliyon
January 2025
Faculty of Medicine and Health Sciences, Sana'a University, Sana'a, Yemen.
Background: Dynamic Coronary Roadmap (DCR) is a new PCI method that may reduce contrast dose and contrast-associated acute kidney injury (CA-AKI) risk. This paper evaluates DCR-guided PCI versus standard angiography PCI for contrast usage, procedure time, and CA-AKI risk.
Methods: On May 1, 2024, we searched PubMed, Scopus, Embase, Cochrane Library, and clinicaltrials.
J Vasc Surg Cases Innov Tech
April 2025
Department of Radiology and Nuclear Medicine, University Teaching and Research Hospital Lucerne, Luzern, Switzerland.
Objective: The aim of this single-center case series is to demonstrate that an ultra-low dose (ULD) can be routinely achieved in the hybrid operating room in standard endovascular aortic repair (EVAR) for infrarenal abdominal aortic aneurysm by adjusting the manufacturer's predefined imaging parameters, hardware configurations and user protocols (including benchmarking).
Methods: The hybrid operating room manufacturer predefined EVAR software setup of the dose exposure control software (OPTIQ, Siemens Healthineers, Forchheim, Germany) at our university medical center was screened for possible improvements regarding radiation dose application. Tests on a water-equivalent as well as polymethyl methacrylate phantom model to assess the impact of technical settings were performed, including comparison of settings for exposure control software, different magnification, collimation configurations and detector distance.
Med Phys
January 2025
Department of Medical Imaging, Radboud University Medical Center, Nijmegen, The Netherlands.
Background: Dedicated breast computed tomography (bCT) systems offer detailed imaging for breast cancer diagnosis and treatment. As new bCT generations are developed, it is important to evaluate their imaging performance and dose efficiency to understand differences over previous models.
Purpose: To characterize the imaging performance and dose efficiency of a second-generation (GEN2) bCT system and compare them to those of a first-generation (GEN1) system.
Am Heart J
January 2025
Department of Cardiology, Columbia University Irving Medical Center/New York-Presbyterian, New York, NY 10032.
Background: This retrospective study addresses the role of operator and fluoroscopy equipment in reducing patient radiation exposure in the Cath lab.
Methods: Data from 99,400 procedures performed in our institution between 2007 and 2019 were reviewed. Dosimetric parameters included reference point air kerma (K), Kerma Area Product (P), fluoroscopic time, and contrast volume.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!