Objective: The Endurant (Medtronic, Minneapolis, Minn) is a new stent graft specifically designed to make more patients anatomically eligible for endovascular aneurysm (EVAR). This study presents the 1-year results of 100 consecutive patients with abdominal aortic aneurysms (AAAs) treated with the Endurant stent graft in real-life practice.
Methods: All clinical preoperative, operative, postoperative, and 1-year follow-up data of patients with the Endurant stent graft from three tertiary centers were prospectively collected. Patients underwent computed tomographic angiography (CTA) preoperatively, at 1 month, and at 1-year post-EVAR. The first 100 patients with an implantation date at least 1 year before our date of analysis and complete information were included. Clinical data, AAA characteristics, presence of endoleaks, graft migration, and other EVAR-related complications were noted. All values are stated as mean ± SD (range).
Results: This study included 100 patients with AAAs (88 men) with a mean age of 73 ± 8 years (47 to 87 years), an AAA size of 61 ± 10 mm (31 to 93 mm), an AAA volume of 210 ± 122 mL (69 to 934 mL), a proximal neck length of 33 ± 14 mm (9 to 82 mm), and an infrarenal angulation of 44 ± 25° (0°-108°). Nineteen of the 100 included patients had at least one anatomic characteristic that was considered a violation of the instructions for use (IFU) of the Endurant stent graft. A primary technical success was achieved in 98% of the patients (one additional stent placement in renal artery was required; one unplanned aorto-uni-iliac device placed), with no primary type I or III endoleaks or conversions. A secondary technical success was achieved in all cases. The 30-day mortality was 2% and the first postoperative CTA documented 16 endoleaks (16%; 16 type II). One-year follow-up showed three iliac limb occlusions (3%), one infected stent graft (causing a type Ia endoleak), and five endovascular reinterventions (5%; three to treat iliac limb occlusions, one proximal extension cuff; and one stent in the renal artery). The 1-year all-cause mortality rate was 12% (12 patients) and the AAA-related mortality was 3%. The mean AAA size was significantly smaller after 1 year (diameter, 54 ± 11.8 [32-80] mm; P < .01; volume, 173 ± 119 [42-1028] mL; P < .01), and one graft migration >5 mm and 13 endoleaks were noted (12 type II, 1 type I [neck dilatation]).
Conclusion: The treatment of patients with AAAs with the Endurant stent graft seems to be successful and durable during the first year after EVAR. Despite the wider inclusion criteria for the Endurant, and with 19% of our patients treated outside the IFU, the AAA-related mortality, number of type I or III endoleaks, and reintervention rates are comparable to the results of other stent grafts.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jvs.2011.02.053 | DOI Listing |
Multimed Man Cardiothorac Surg
January 2025
Congenital Heart Center, Division of Cardiovascular Surgery, Department of Surgery, University of Florida, Gainesville, FL, USA.
The Berlin Heart EXCOR is a pulsatile paracorporeal ventricular assist device (VAD) for neonates, infants, children and adults with congenital or acquired severe ventricular dysfunction. Berlin Heart EXCOR VADs are routinely used as either a bridge to a cardiac transplantation, or occasionally as a bridge to ventricular recovery. Our programmatic philosophy is to bridge neonates and infants with functionally univentricular ductal-dependent systemic circulation or functionally univentricular ductal-dependent pulmonary circulation who are at high risk for staged palliation because of important cardiac risk factors with a single-ventricle VAD (sVAD) as a bridge to a cardiac transplant.
View Article and Find Full Text PDFAnn Vasc Surg
January 2025
Department of for Cardiovascular Surgery, University Heart Centre Freiburg - Bad Krozingen, University Medical Centre Freiburg, Freiburg, Germany; Faculty of Medicine, Albert Ludwigs University Freiburg, Freiburg, Germany.
Objectives: To report outcomes after the use of the Omniflow II biosynthetic graft (LeMaitre Vascular, Il, USA) for vascular reconstruction in patients with prosthetic infection at the aorto-iliac and femoropopliteal level.
Methods: Within a six-year period, all consecutive patients with aorto-iliac and femoro-popliteal graft infection treated by resection of the infected graft material, extensive local debridement and reconstruction using Omniflow II biosynthetic graft were retrospectively analzyed. Patient characteristics, intraoperative details, postoperative outcomes, and infection details were assessed.
Eur J Vasc Endovasc Surg
January 2025
Aortic Center, Hôpital Marie Lannelongue, GHPSJ, Le Plessis-Robinson, France. Electronic address:
Objective: The aim of this study was to perform a comparative pre-clinical evaluation of a new fenestrated endovascular aneurysm repair dedicated stent using perfused 3D printed patient anatomies.
Methods: The test setup included a pulsatile pump set to reproduce human haemodynamics, four 3D printed pararenal aneurysms connected to a bench test, and four corresponding fenestrated grafts (Zenith Fenestrated; Cook, Bloomington, IN, USA). Bridging stents were sized based on analysis of patients computed tomography scans and included either four standard covered stents (BeGraft [BG]; Bentley, Hechingen, Germany) or four similar covered stent grafts mounted on a modified balloon to allow stent implantation and flaring in one step (BeFlared [BF]; Bentley).
J Cardiovasc Surg (Torino)
February 2025
Department of Vascular Surgery, AZ Sint Blasius, Dendermonde, Belgium.
Background: Due to the heterogeneity of literature findings, stent type selection for the endovascular treatment of complex aorto-iliac occlusive disease remains challenging. The BELSTREAM study, a physician-initiated, prospective, multicenter, single-arm study, aims to report the safety and efficacy of the balloon expandable LifeStream Peripheral Stent Graft System (BD, Tempe, Arizona, USA) for the treatment of complex TASC C and D aorto-iliac artery lesions.
Methods: Seventy patients and 133 lesions were included at six Belgian institutions.
Ann Vasc Dis
January 2025
Department of Vascular Surgery, The University of Tokyo, Tokyo, Japan.