Background: Diagnostic laboratories need automation that facilitates efficient processing and workflow management to meet today's challenges for expanding services and reducing cost, yet maintaining the highest levels of quality.

Methods: Processing efficiency of two commercially available automated systems for quantifying HIV-1 and HCV RNA, Abbott m2000 system and Roche COBAS Ampliprep/COBAS TaqMan 96 (docked) systems (CAP/CTM), was evaluated in a mid/high throughput workflow laboratory using a representative daily workload of 24 HCV and 72 HIV samples. Three test scenarios were evaluated: A) one run with four batches on the CAP/CTM system, B) two runs on the Abbott m2000 and C) one run using the Abbott m2000 maxCycle feature (maxCycle) for co-processing these assays. Cycle times for processing, throughput and hands-on time were evaluated.

Results: Overall processing cycle time was 10.3, 9.1 and 7.6 h for Scenarios A), B) and C), respectively. Total hands-on time for each scenario was, in order, 100.0 (A), 90.3 (B) and 61.4 min (C).

Conclusions: The interface of an automated analyzer to the laboratory workflow, notably system set up for samples and reagents and clean up functions, are as important as the automation capability of the analyzer for the overall impact to processing efficiency and operator hands-on time.

Download full-text PDF

Source
http://dx.doi.org/10.1515/CCLM.2011.625DOI Listing

Publication Analysis

Top Keywords

abbott m2000
16
hands-on time
12
roche cobas
8
cobas ampliprep/cobas
8
ampliprep/cobas taqman
8
hcv rna
8
processing efficiency
8
processing
5
improving clinical
4
clinical laboratory
4

Similar Publications

Article Synopsis
  • * Involving 200 newly diagnosed HIV-1 patients, the research revealed significant correlations, such as lower hemoglobin levels linked to higher CD4+ counts and lower PVL, while abnormalities in various blood parameters were more pronounced in those co-infected with HIV-TB.
  • * The findings suggest that specific hematological and biochemical indicators, including hemoglobin and inflammatory markers, could serve as early warning signs for TB in HIV-infected individuals, enhancing patient management
View Article and Find Full Text PDF

Introduction Anti-tumor necrosis factor alpha (TNFα) therapies have revolutionized the management of Crohn's disease (CD). However, they increase the risk of viral reactivation, particularly hepatitis B virus (HBV). This study aims to define the HBV serological profiles of patients with CD who are candidates for biological therapy, identifying profiles at potential risk for reactivation or exacerbation following immunosuppressive treatment.

View Article and Find Full Text PDF

Introduction: HCV antibody assays have been used to screen for HCV, but confirmation of acute infection is dependent on RNA or core antigen testing. The aim of the study was to compare the performance of five HCV test methods, including RNA testing, on a US emergency department population.

Methods: Clinical performance metrics were calculated on 708 consenting Johns Hopkins Emergency Department patients who self-reported an increased risk for HCV infection.

View Article and Find Full Text PDF

Residual transmission of HIV infection from mother to child in the Atlantic and littoral departments in Benin.

BMC Pediatr

October 2024

Laboratory of Biology and Molecular Typing in Microbiology (LBTMM), Faculty of Sciences and Technology (FAST), University of Abomey - Calavi, Cotonou, 05 BP 1604, Benin.

Article Synopsis
  • The study evaluates the effectiveness of WHO's recommendations in Benin regarding the prevention of HIV transmission to exposed infants, using data from 2016 to 2021.* -
  • Analysis involved 5,312 infants, revealing a 3.4% overall HIV transmission rate, impacted by factors such as maternal treatment methods and feeding practices, with lower transmission linked to mothers on tritherapy.* -
  • The findings highlight the success of the PMTCT program in reducing HIV transmission among infants, emphasizing the importance of consistent treatment for mothers of childbearing age.*
View Article and Find Full Text PDF

Objectives: We sought to evaluate the analytical performance of the Alinity m system (Abbott Molecular) and to compare the clinical performance of HIV-1 assays on the Alinity m and m2000 RealTime platforms (Abbott Molecular).

Methods: The sensitivity, precision, and accuracy of the Alinity m instrument were determined using a panel of standard samples (n = 46). The carryover effect was assessed by analyzing HIV-negative clinical samples (n = 20).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!