Solvent microextraction into a single drop.

Anal Chem

Department of Chemistry, University of Alberta, Edmonton, Alberta, Canada T6G 2G2.

Published: July 1996

An analytical technique is described which combines solvent extraction with gas chromatographic (GC) analysis in a simple and inexpensive apparatus involving very little solvent consumption. A small drop (8 μL) of a water-immiscible organic solvent, containing an internal standard, is located at the end of a Teflon rod which is immersed in a stirred aqueous sample solution. After the solution has been stirred for a prescribed period of time, the probe is withdrawn from the aqueous solution, and the organic phase is sampled with a microsyringe and injected into the GC for quantification. The observed rate of solvent extraction is in good agreement with a convective-diffusive kinetic model. Analytically, the relative standard deviation of the method is 1.7% for a 5.00-min extraction of the analyte 4-methylacetophenone into n-octane.

Download full-text PDF

Source
http://dx.doi.org/10.1021/ac960042zDOI Listing