Purpose: To investigate the accuracy and potential limitations of MV image-based dynamic multileaf collimator (DMLC) tracking in a porcine model on a linear accelerator.

Methods And Materials: A thermo-expandable NiTi stent designed for kilovoltage (kV) X-ray visualization of lung lesions was inserted into the bronchia of three anaesthetized Göttingen minipigs. A four-dimensional computed tomography scan was used for planning a five-field conformal treatment with circular multileaf collimator (MLC) apertures. A 22.5 Gy single fraction treatment was delivered to the pigs. The peak-to-peak stent motion was 3 to 8 mm, with breathing periods of 1.2 to 4 s. Before treatment, X-ray images were used for image-guided setup based on the stent. During treatment delivery, continuous megavoltage (MV) portal images were acquired at 7.5 Hz. The stent was segmented in the images and used for continuous adaptation of the MLC aperture. Offline, the tracking error in beam's eye view of the treatment beam was calculated for each MV image as the difference between the MLC aperture center and the segmented stent position. The standard deviations of the systematic error Σ and the random error σ were determined and compared with the would-be errors for a nontracking treatment with pretreatment image-guided setup.

Results: Reliable stent segmentation was obtained for 11 of 15 fields. Segmentation failures occurred when image contrast was dominated by overlapping anatomical structures (ribs, diaphragm) rather than by the stent, which was designed for kV rather than MV X-ray visibility. For the 11 fields with reliable segmentation, Σ was 0.5 mm/0.4 mm in the two imager directions, whereas σ was 0.5 mm/1.1 mm. Without tracking, Σ and σ would have been 1.7 mm/1.4 mm and 0.8 mm/1.4 mm, respectively.

Conclusion: For the first time, in vivo DMLC tracking has been demonstrated on a linear accelerator showing the potential for improved targeting accuracy. The study mimicked the envisioned patient workflow of future patient treatments. Clinical implementation of MV image-based tracking would require markers designed for MV visibility.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3162131PMC
http://dx.doi.org/10.1016/j.ijrobp.2011.03.023DOI Listing

Publication Analysis

Top Keywords

multileaf collimator
12
image-based dynamic
8
dynamic multileaf
8
stent
8
niti stent
8
linear accelerator
8
dmlc tracking
8
stent designed
8
mlc aperture
8
tracking
6

Similar Publications

Purpose: This study presents novel quality assurance (QA) approach for volumetric modulated arc therapy (VMAT) that leverages frame-by-frame electronic portal imaging device (EPID) images integrated into Mobius3D for accurate three-dimensional dose calculations.

Methods: Sequential EPID images for VMAT plans were acquired every 0.4-second by iView system and processed through iterative deconvolution to mitigate blurring from photon scattering.

View Article and Find Full Text PDF

Design and performance of a novel multi-leaf collimator composed of leaves with fixed and movable layers.

Med Phys

November 2024

Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.

Background: The multi-leaf collimator (MLC) is an advanced device utilized for beam shaping and intensity modulation in radiotherapy. With the framework of the contemporary single-layer MLC featuring a rounded leaf tip, the leaf tip transmission and leakage exert a considerable influence on radiotherapy.

Purpose: To scale down the leakage and transmission from the leaf gap when the opposite leaves are closed.

View Article and Find Full Text PDF

Novel application of an adapted virtual cone technique for the radiosurgical treatment of a 3 mm brain metastasis with a standard multileaf collimator.

Med Dosim

November 2024

Ascension NE Wisconsin St Elizabeth Campus, Appleton, WI 54915, USA. Electronic address:

The virtual cone is an innovative MLC-based technique for generating dose distributions comparable to those of physical stereotactic cones. Initially designed for functional radiosurgery applications using a high-definition multileaf collimator (MLC) with 2.5 mm leaf width, this technique has been adapted to a standard 5 mm MLC system for treating small brain metastases.

View Article and Find Full Text PDF

Dosimetric effects of the leaf positioning error of the halcyon(2.0) dual-layer multileaf collimator (MLC) on the rectal cancer radiotherapy.

Med Dosim

November 2024

Department of Radiation Oncology, The Affiliated Cancer Hospital of Nanjing Medical University, Jiangsu Cancer Hospital, Jiangsu Institute of Cancer Research, Nanjing, 210009, China; Department of Environmental Genomics, Jiangsu Key Laboratory of Cancer Biomarkers, Prevention and Treatment, Collaborative Innovation Center for Cancer Personalized Medicine, Nanjing Medical University, Nanjing, 210024, China. Electronic address:

To study dosimetric effects of leaf positioning errors (LPEs) of the Halcyon(2.0) dual-layer MLC on the long-course chemoradiotherapy (LCCRT) with 45∼50.4 Gy in 25∼28 fractions for rectal cancer.

View Article and Find Full Text PDF

This study aimed to comprehensively investigate the influence of multi-leaf collimator (MLC) position errors on both the clinical absolute dose distribution and Gamma passing rate (%GP) in intensity-modulated radiation therapy (IMRT) plans for breast cancer. Additionally, the correlation between %GP and the clinical absolute dose relative difference (%DE) caused by MLC position errors was analysed. Ten IMRT plans for breast cancer were randomly selected.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!