Background: Electromagnetic interference (EMI) induced by electrocautery during surgery in patients with cardiac pacemakers or implanted cardioverter-defibrillators (ICDs) may inhibit pacing and cause inappropriate tachyarrhythmia oversensing. In particular, susceptibility to EMI may be enhanced in ICDs by frequently used wide interelectrode sensing (i.e., integrated bipolar sensing). Consequently, ICD function is usually disabled preoperatively and restored later by noninvasive programming. Because sensing by closely spaced electrodes (i.e., true bipolar) may be less susceptible to EMI, preoperative programming to a true bipolar mode may minimize the need for perioperative programming while preserving device function.

Methods: Our study population consisted of 23 consecutive patients either receiving a new ICD or undergoing ICD pulse generator change. In each patient, electrocautery-induced EMI was initiated with the ICD in the closely spaced sensing configuration and again during widely spaced sensing.

Results: In comparing the 2 sensing modes, right ventricular electrogram amplitude was significantly greater and EMI noise amplitude tended to be greater with widely spaced bipolar sensing. Furthermore, widely spaced bipolar sensing was associated with ICD pacing inhibition in 22 of 23 patients and incorrect "ventricular fibrillation" detection in 17 of 23 patients. Conversely, closely spaced bipolar sensing was not accompanied by either pacing inhibition or incorrect ventricular fibrillation sensing.

Conclusion: Closely spaced bipolar sensing (i.e., true bipolar) appropriately rejects electrocautery-induced EMI. Programming implanted devices to closely spaced bipolar sensing may minimize the need for perioperative reprogramming while preserving intraoperative device operation.

Download full-text PDF

Source
http://dx.doi.org/10.1213/ANE.0b013e3182172a18DOI Listing

Publication Analysis

Top Keywords

bipolar sensing
28
closely spaced
24
spaced bipolar
24
true bipolar
12
sensing
11
bipolar
10
spaced
9
electromagnetic interference
8
devices closely
8
minimize perioperative
8

Similar Publications

The application of the technique for dorsal median sulcus mapping in intramedullary space occupying surgery: a single-center experience.

Acta Neurochir (Wien)

January 2025

Department of Neurosurgery and Department of Neuroscience, Fujian Key Laboratory of Brain Tumors Diagnosis and Precision Treatment, Xiamen Key Laboratory of Brain Center, the First Affiliated Hospital of Xiamen University, School of Medicine, Xiamen University, Xiamen, Fujian, China.

Purpose: To investigate the technique for dorsal median sulcus (DMS) mapping and assess its application value in preserving dorsal columnn (DC) function during intramedullary space occupying surgery based on a single-center experience.

Methods: A retrospective analysis was conducted on 41 cases of intramedullary spinal cord tumor admitted to the Department of Neurosurgery at the First Affiliated Hospital of Xiamen University from March 2017 to August 2023. All included cases underwent intraoperative electrophysiological monitoring, and were divided into a study group (n = 18) and a control group (n = 23), based on whether DMS mapping technique was utilized.

View Article and Find Full Text PDF

Local field potential (LFP) recordings using chronically implanted sensing-enabled stimulators are a powerful tool for indexing symptom presence and severity in neurological and neuropsychiatric disorders, and for enhancing our neurophysiological understanding of brain processes. LFPs have gained interest as input signals for closed-loop deep brain stimulation (DBS) and can be used to inform DBS parameter selection. LFP recordings using chronically implanted sensing-enabled stimulators have various implementational challenges.

View Article and Find Full Text PDF

5-Hydroxyindoleacetic acid (5-HIAA), a vital metabolite of serotonin (5-HT), is crucial for understanding metabolic pathways and is implicated in various mental disorders. In situ monitoring of 5-HIAA is challenging due to the lack of affinity ligands and issues with electrochemical fouling. We present an advanced sensing approach that integrates customizable molecular imprinting polymer (MIP) with self-driven galvanic redox potentiometry (GRP) for precise, real-time in vivo monitoring of 5-HIAA.

View Article and Find Full Text PDF

Characteristics of In Vivo Lesion Formation With a Temperature-Controlled Diamond-Tip Radiofrequency Ablation Catheter in the Ventricle: A Preclinical Model.

Circ Arrhythm Electrophysiol

January 2025

Department of Cardiovascular Medicine, Mayo Clinic, Rochester, MN (T.H., M.E.R., O.Y., G.N.K., N.O., T.K., L.N., D.L.P., K.C.S.).

Background: Power-controlled radiofrequency ablation with irrigated-tip catheters has been the norm for ventricular ablation for almost 2 decades. New catheter technology has recently integrated more accurate tissue temperature sensing enabling temperature-controlled irrigated ablation. We aimed to investigate the in vivo ablation parameters and lesion formation characteristics in ventricular myocardium using a novel temperature-controlled radiofrequency catheter.

View Article and Find Full Text PDF

Optoelectronic synapse devices (OESDs) inspired by human visual systems enable to integration of light sensing, memory, and computing functions, greatly promoting the development of in-sensor computing techniques. Herein, dual-mode integration of bipolar response photodetectors (PDs) and artificial optoelectronic synapses based on ZnO/SnSe heterojunctions are presented. The function of the fabricated device can be converted between the PDs and OESDs by modulating the light intensity.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!