Purpose: To develop a new rabbit model of arterial stenosis using endovascular radiofrequency (RF) energy.

Materials And Methods: Ten rabbits were used for multiple endovascular RF applications to the aorta and left common carotid artery through the Habib™ VesCoag™ catheter. Angiography and color Doppler ultrasound were used to assess vessel patency immediately following the procedure and six weeks later. One rabbit was sacrificed following the procedure for histopathologic analysis of the vessel wall. Two rabbits died of aortic and carotid rupture, respectively, immediately after the procedure. The remaining seven rabbits were sacrificed after six-week follow-up for histopathological analysis.

Results: Optimal RF generator settings to induce significant arterial stenosis (>50%) without complications were standardized at 24-26 watts (W) for 1.5 min for the aorta and 6 W for 1 min for the common carotid artery. The six-week follow-up showed permanent results in all surviving rabbits. Histopathology revealed intima and medial smooth muscle layer necrosis.

Conclusion: We have developed a novel rabbit model of arterial stenosis using endovascular RF energy. Our model is fast, safe, inexpensive, and reproducible. It would be useful for experimental investigations and new therapeutic devices.

Download full-text PDF

Source
http://dx.doi.org/10.3109/08941939.2011.557470DOI Listing

Publication Analysis

Top Keywords

arterial stenosis
16
model arterial
12
endovascular radiofrequency
8
rabbit model
8
stenosis endovascular
8
common carotid
8
carotid artery
8
six-week follow-up
8
novel experimental
4
experimental animal
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!