Purpose: Clinicians often do not have the benefit of adequate safety or clinical data when evaluating the merit of either newly marketed implant devices or novel clinical procedures. This has been the case for dental implants following the initial documentation of their safety and efficacy and is demonstrated in the evolution of immediate load application. Following demonstration of safety and successful application of an implant in an animal study prior to its market release, this report provides the clinical outcomes for the first 100 Ti-Unite implants provided to 24 patients in a clinical practice over 9 years.

Materials And Methods: An electronic record/clinical database review of consecutive early loaded implants from a multiple surgeon/single prosthodontist practice was conducted for quality assurance. Data extraction of standard exposure and outcome variables was accomplished by a trained individual not affiliated with the clinical practice.

Results: The results revealed one failure before and none following definitive restoration with a variety of prostheses. The mean length of time from immediate to definitive restorations was 5.3 ± 1.1 months for crowns, 3.9 ± 1.3 months for fixed partial dentures, and 7.8 ± 4.1 months for mandibular "hybrid" prostheses. The most common unexpected findings during the initial three postinsertion visits were lost access restoration and cement failure.

Conclusions: Pre-market animal data regarding the safety and success of a new implant used with an early loading protocol was replicated in the clinical results of the first 100 implants used in practice. Additionally, the clinical results are favorable when compared to conventional loading protocols from this same practice and provide helpful comparative metrics (delayed vs immediate loading) to use when discussing implant treatment with patients.

Download full-text PDF

Source

Publication Analysis

Top Keywords

ti-unite implants
8
animal study
8
clinical
7
implants
5
restoration ti-unite
4
implants practice-based
4
practice-based evidence
4
evidence compared
4
compared animal
4
study outcomes
4

Similar Publications

To investigate if it is possible to achieve complete decontamination of dental implant surfaces with different surface characteristics. Twelve implant pieces with an Osseotite surface and 12 implant pieces with a Ti-Unite surface were attached on to the complete lower dentures of six patients and were allowed to accumulate plaque for 30 days. When retrieved, the implant decontamination protocol used, involved both mechanical (PeriBrush™) and chemical (3% HO) decontamination.

View Article and Find Full Text PDF

The success of dental implants depends mainly on osseointegration and gingival sealing. Therefore, early attachment and spreading of epithelial cells might be critical for a positive outcome. Research in dental implant materials has primarily focused on surface roughness, defined by the average roughness (Ra) index, as it promotes the process of osseointegration.

View Article and Find Full Text PDF

Purpose: Clinicians often do not have the benefit of adequate safety or clinical data when evaluating the merit of either newly marketed implant devices or novel clinical procedures. This has been the case for dental implants following the initial documentation of their safety and efficacy and is demonstrated in the evolution of immediate load application. Following demonstration of safety and successful application of an implant in an animal study prior to its market release, this report provides the clinical outcomes for the first 100 Ti-Unite implants provided to 24 patients in a clinical practice over 9 years.

View Article and Find Full Text PDF

Aims: This study was designed to evaluate the effect of gap width and graft placement on bone healing around implants placed into simulated extraction sockets in the mandibles of four beagle dogs.

Materials And Methods: Four Ti-Unite implants (13 mm x 3.3 mm) were placed on each side of the mandible.

View Article and Find Full Text PDF

Objective: The purpose of the study was to investigate whether osseointegration can occur on rough implant surfaces that previously had been coated with bacterial biofilm.

Materials And Methods: The premolars on both sides of the mandible in four beagle dogs were extracted. Following 3 months healing, three titanium implants Ti-Unite, Nobel Biocare were partially inserted in the left side of each mandible.