Background: The BacT/ALERT microbial detection system (BTA) is used for testing leukoreduced apheresis platelets (LR-AP) in plasma. Platelet additive solutions (PASs) such as InterSol (PAS III) may be used to reduce the amount of plasma transfused in LR-AP. This study evaluated the performance of the two-bottle BTA testing scheme in the recovery of seeded microorganisms from LR-AP in InterSol-plasma compared to a reference plate culture method.

Study Design And Methods: Hyperconcentrated, double LR-AP were collected from healthy donors; InterSol was added (65% Intersol:35% plasma), equally divided into two containers, and then inoculated with an isolate of 1 of 10 clinically relevant index organisms at two levels. Aerobic (BPA) and anaerobic (BPN) BTA bottles were inoculated with 4 mL each of the inoculated LR-AP, and blood agar plates (BAPs) for aerobic and anaerobic culture (0.5 mL each).

Results: Zero false-positives from 103 bottle pairs were observed. All 400 two-bottle BTA tests were positive within 24 hours, except for Propionibacterium acnes (maximum time-to-detection of 86.4 hr) and 13 of 20 pairs of Streptococcus viridans (maximum time-to-detection of 31.7 hr). Thirteen of 400 BAP two-plate tests were negative for starting bacterial concentrations of 10 colony-forming units (CFUs)/mL or less. At 40 CFUs/mL or less, BTA was 100% positive while BAP was 94% positive.

Conclusion: Seeded organism recovery was superior in the two-bottle BTA test system compared to the two-plate BAP system using InterSol platelets (PLTs). This performance is comparable to previously published results for PLTs in plasma. The use of InterSol does not appear to have a detrimental effect on the performance of the two-bottle BTA system.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1537-2995.2011.03149.xDOI Listing

Publication Analysis

Top Keywords

two-bottle bta
16
microbial detection
8
detection system
8
pas iii
8
platelet additive
8
bta testing
8
performance two-bottle
8
maximum time-to-detection
8
bta
7
system
5

Similar Publications

Background: The BacT/ALERT microbial detection system (BTA) is used for testing leukoreduced apheresis platelets (LR-AP) in plasma. Platelet additive solutions (PASs) such as InterSol (PAS III) may be used to reduce the amount of plasma transfused in LR-AP. This study evaluated the performance of the two-bottle BTA testing scheme in the recovery of seeded microorganisms from LR-AP in InterSol-plasma compared to a reference plate culture method.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!