Magnetic Properties of Nanoparticles Useful for SQUID Relaxometry in Biomedical Applications.

J Magn Magn Mater

Senior Scientific LLC, 11109 Country Club NE, Albuquerque, New Mexico, 87111.

Published: March 2011

We use dynamic susceptometry measurements to extract semiempirical temperature-dependent, 255 to 400 K, magnetic parameters that determine the behavior of single-core nanoparticles useful for SQUID relaxometry in biomedical applications. Volume susceptibility measurements were made in 5K degree steps at nine frequencies in the 0.1 - 1000 Hz range, with a 0.2 mT amplitude probe field. The saturation magnetization (M(s)) and anisotropy energy density (K) derived from the fitting of theoretical susceptibility to the measurements both increase with decreasing temperature; good agreement between the parameter values derived separately from the real and imaginary components is obtained. Characterization of the Néel relaxation time indicates that the conventional prefactor, 0.1 ns, is an upper limit, strongly correlated with the anisotropy energy density. This prefactor decreases substantially for lower temperatures, as K increases. We find, using the values of the parameters determined from the real part of the susceptibility measurements at 300 K, that SQUID relaxometry measurements of relaxation and excitation curves on the same sample are well described.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3079235PMC
http://dx.doi.org/10.1016/j.jmmm.2010.10.042DOI Listing

Publication Analysis

Top Keywords

squid relaxometry
12
susceptibility measurements
12
nanoparticles squid
8
relaxometry biomedical
8
biomedical applications
8
anisotropy energy
8
energy density
8
measurements
5
magnetic properties
4
properties nanoparticles
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!