Purpose: Patients undergoing left atrial ablation require transseptal puncture, which can be challenging, even for experienced physicians. This study compared the efficacy and safety of radiofrequency (RF) energy transseptal punctures to conventional approach.

Methods: Patients requiring transseptal puncture for left atrial access were included using either conventional approach or the NRG™ RF transseptal needle as first attempt. Procedure time for transseptal access, fluoroscopy time, crossover, and safety of both techniques were compared.

Results: A total of 241 transseptal punctures were performed in 148 consecutive patients (114 men, 54 ± 10 years, left atrial volume 32 ± 10 ml/m(2)) who underwent 157 procedures with left atrial access, mainly for atrial fibrillation. It was a repeat transseptal procedure in 49 patients. Procedures were guided by transesophageal echocardiography. RF transseptal puncture was planned in 119 procedures. RF was delivered in 98 procedures (82%) for 139/187 punctures: 48 punctures did not require RF, including 25 punctures performed by exposing the needle tip, 22 through patent foramen ovale, and 1 RF delivery failure by the generator. Average time for RF transseptal was 4.8 ± 2.8 min compared to 7.5 ± 8.5 min for conventional approach (p = 0.045). Fluoroscopy time was 1.8 ± 1.3 min for RF transseptal and 2.9 ± 2.8 min for standard approach (p = 0.043). Four patients required crossover to RF transseptal needle in the conventional group (p = 0.003). One tamponade occurred at the end of procedure in a patient using the RF needle, and one interatrial septum dissection with aortic root hematoma occurred in the conventional group.

Conclusions: Transseptal needle puncture using RF energy can be performed safely and quickly under imaging guidance.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10840-011-9564-2DOI Listing

Publication Analysis

Top Keywords

transseptal puncture
16
transseptal needle
16
left atrial
16
transseptal
14
needle puncture
8
radiofrequency energy
8
transseptal punctures
8
atrial access
8
conventional approach
8
time transseptal
8

Similar Publications

Transjugular transcatheter edge-to-edge mitral valve repair in a patient with functional mitral regurgitation: a case report.

Eur Heart J Case Rep

January 2025

Department of Cardiovascular Medicine, Sendai Kousei Hospital, 1-20 Tsutsumidori-amamiya, Aoba Ward, Sendai, Miyagi 9810914, Japan.

Background: Transcatheter edge-to-edge mitral valve repair (M-TEER) using the MitraClip system is primarily performed using the transfemoral approach. However, when this approach is not feasible, the transjugular approach can be used as an alternative.

Case Summary: A 57-year-old man presented with heart failure and persistent New York Heart Association class IV symptoms, refractory to guideline-directed medical therapy, intravenous therapy, and intra-aortic balloon pumping.

View Article and Find Full Text PDF

Intracardiac echocardiography (ICE) has been used to guide radio-frequency catheter ablation (RFCA) for better catheter navigation and less radiation exposure in treating atrial fibrillation (AF). This retrospective cohort study enrolled 227 AF patients undergoing ICE- or traditional fluoroscopy (TF)-guided RFCA for AF in a tertiary hospital. ICE was used more often in patients with atrial tachycardia [odds ratio (OR) 3.

View Article and Find Full Text PDF

Novel algorithm for non-invasive estimation of left atrial pressure in patients with atrial fibrillation.

Eur Heart J Cardiovasc Imaging

December 2024

Division of Cardiology, Severance Cardiovascular Hospital, Yonsei University College of Medicine, 50-1 Yonsei-ro, Seodaemun-gu, Seoul 03722, Republic of Korea.

Aims: Determining elevated left atrial (LA) pressure is crucial in patients with atrial fibrillation (AF), yet non-invasive estimation using echocardiography remains unclear. This study aimed to identify useful echocardiographic indices for identifying elevated LA pressure in patients with AF.

Methods And Results: Patients with paroxysmal or persistent AF referred for catheter ablation at two tertiary hospitals were prospectively enrolled.

View Article and Find Full Text PDF

Case Report: A novel method of needle-free transseptal puncture.

Front Cardiovasc Med

December 2024

Department of Internal Medicine, College of Medicine, National Taiwan University, Taipei, Taiwan.

Background: In the era of fluoroless catheter ablation (CA), achieving a successful transseptal puncture (TSP) presents a significant challenge. We introduce a novel technique for zero-fluoroscopy and cost-effective needle-free TSP.

Case Summary: We describe two cases where a GMS-1 guidewire (0.

View Article and Find Full Text PDF

Background: The closure of a patent foramen ovale (PFO) using transseptal puncture has particular advantages and disadvantages. Thus, transseptal puncture should be re-evaluated in detail.

Aims: We aimed to assess the effectiveness of the detailed transseptal puncture technique in patients who underwent PFO closure due to cryptogenic stroke or transient ischemic attack in terms of residual shunts and atrial fibrillation.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?