This work presents the use of a flow injection capacitive immunosensor to detect staphylococcal enterotoxin A (SEA). The study was based on the direct detection of a capacitance change due to the binding between SEA and anti-SEA immobilized on a gold electrode. The optimal regeneration solution, flow rate, sample volume and buffer conditions were studied. Under the optimum conditions, this label-free biosensor provided linearity between 1 × 10(-12) g L(-1) and 1 × 10(-8) g L(-1) of SEA and the limit of detection was 1 × 10(-12) g L(-1) which was much lower than the infectious dose (0.5 × 10(-6) - 1 × 10(-6) g L(-1)). Using the regeneration solution of, 15.0 mM glycine-HCl pH 2.20, to break the binding between SEA and the immobilized anti-SEA enabled the electrode to be reused up to 39 times. This technique was applied to analyze SEA in liquid and solid food samples. Any matrix effect can be eliminated by simple dilution. SEA contamination was found in three samples, iced tea with milk (28 ± 1 ng L(-1)), orange juice (113 ± 6 ng L(-1)) and fried chicken (1.1 ± 0.2 ng g(-1)); however, the concentrations were much lower than the infectious dose. The proposed method would be useful for rapid screening of SEA in various matrices.

Download full-text PDF

Source
http://dx.doi.org/10.1080/10934529.2011.562810DOI Listing

Publication Analysis

Top Keywords

staphylococcal enterotoxin
8
sea
8
enterotoxin sea
8
capacitive immunosensor
8
binding sea
8
regeneration solution
8
10-12 l-1
8
lower infectious
8
infectious dose
8
l-1
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!