Background: Little is known regarding image quality and the required radiation dose for step-and-shoot and retrospective coronary computed tomography angiography (CCTA) with tube current modulation (TCM) in 128-slice multidetector CT (MDCT) coronary angiography.

Purpose: To compare image quality and radiation dose in patients who underwent 128-slice MDCT by the step-and-shoot method with those in patients who underwent 128-slice MDCT with retrospective CCTA with TCM.

Material And Methods: CCTA obtained with 128-slice MDCT was retrospectively evaluated in 160 patients. Two independent reviewers separately scored the subjective image quality of the coronary artery segments (1, excellent; 4, poor) for step-and-shoot (68, mean heart rate [HR]: 59.3 ± 6.8) and retrospective CCTA with TCM (77, mean HR: 59.1 ± 9.8). Interobserver variability was calculated. Effective radiation doses of both scan techniques were calculated with dose-length product.

Results: There was good agreement for quality scores of coronary artery segment images between the independent reviewers (κ = 0.72). The number of coronary artery segments that could not be evaluated was 2.85% (27 of 947) in the step-and-shoot and 1.87% (20 of 1071) in retrospective CCTA with TCM. Image quality scores were not significantly different (P > .05). Mean patient radiation dose was 63% lower for step-and-shoot (1.94 ± 0.70 mSv) than for retrospective CCTA with TCM (4.51 ± 1.18 mSv) (P < 0.0001). For patients who underwent step-and-shoot or retrospective CCTA with TCM, an average HR of 63.5 beats per minute was identified as the threshold for the prediction of non-diagnostic image quality for both protocols. There were no significant differences in the image quality of both methods between obese (body mass index [BMI] ≥ 25) and non-obese patients (BMI < 25), but radiation doses were higher in the obesity group than in the non-obesity group for both methods.

Conclusion: Both step-and-shoot and retrospective CCTA with TCM using 128-slice MDCT had similar subjective image quality scores, but step-and-shoot required a lower radiation dose than retrospective CCTA with TCM.

Download full-text PDF

Source
http://dx.doi.org/10.1258/ar.2010.100290DOI Listing

Publication Analysis

Top Keywords

image quality
32
retrospective ccta
28
ccta tcm
24
128-slice mdct
20
radiation dose
20
step-and-shoot retrospective
12
patients underwent
12
coronary artery
12
quality scores
12
step-and-shoot
9

Similar Publications

Importance: In the Atrial Cardiopathy and Antithrombotic Drugs in Prevention After Cryptogenic Stroke (ARCADIA) randomized clinical trial, anticoagulation did not prevent recurrent stroke among patients with a recent cryptogenic stroke and atrial cardiopathy. It is unknown whether anticoagulation prevents covert infarcts in this population.

Objective: To test the use of apixaban vs aspirin for prevention of nonlacunar covert infarcts after cryptogenic stroke in patients with atrial cardiopathy.

View Article and Find Full Text PDF

Fast, Present and Future of the Concept of Spondyloarthritis.

Curr Rheumatol Rep

January 2025

Rheumatologisches Versorgungszentrum Steglitz, Ruhr Universität Bochum, Schloßstr.110, 12163, Berlin, Germany.

Purpose Of Review: Axial spondyloarthritis (axSpA) is a rather prevalent chronic inflammatory rheumatic disease that affects already relatively young patients. It has been known better since the end of the nineteenth century but quite a lot has been learned since the early 60ies when the first classification (diagnostic) criteria for ankylosing spondylitis (AS) were agreed on. I have been part of many developments in the last 30 years, and I'm happy to have been able to contribute to the scientific progress in terms of diagnosis, imaging, pathophysiology and therapy.

View Article and Find Full Text PDF

[Indications for surgical treatment of traumatic fractures of the thoracic spine and lumbar spine].

Unfallchirurgie (Heidelb)

January 2025

Klinik für Orthopädie, Unfallchirurgie und Plastische Chirurgie, Universitätsklinikum Leipzig AöR, Liebigstraße 20, 04103, Leipzig, Deutschland.

Fractures of the thoracic (Th) and lumbar (L) vertebrae are among the most frequent fracture entities in Germany and particularly affect the thoracolumbar junction (TLJ; Th11-L2). Based on expert recommendations and consensus meetings, the thoracolumbar AOSpine injury score was established for patients with healthy bone and the osteoporotic fracture (OF) score for geriatric patients with the respective classifications for treatment decisions. In both cohorts, the treatment decision is based on the fracture morphology, neurological status and patient-specific contextual factors.

View Article and Find Full Text PDF

Purpose: The depth within the body, small diameter, long length, and varying tissue surrounding the spinal cord impose specific considerations when designing radiofrequency coils. The optimal coil configuration for 7 T cervical spinal cord MRI is unknown and, currently, there are very few coil options. The purpose of this work was (1) to establish a quality control protocol for evaluating 7 T cervical spinal cord coils and (2) to use that protocol to evaluate the performance of 4 different coil designs.

View Article and Find Full Text PDF

The quality of CT images obtained from hepatocellular carcinoma (HCC) patients is complex, affecting diagnostic accuracy, precision, and radiation dose assessment due to increased exposure risks. The study evaluated image quality qualitatively and quantitatively by comparing quality levels with an effective radiation dose to ensure acceptable quality accuracy. This study retrospectively reviewed 100 known HCC patients (Li-RADS-5) who underwent multidetector computed tomography (MDCT) multiphasic scans for follow-up of their health condition between January and October 2023.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!