Randomized clinical trial comparing double-lumen tube and EZ-Blocker for single-lung ventilation.

Br J Anaesth

Department of Cardiothoracic and Vascular Anesthesia and Intensive Care Medicine, University of Vienna, Waehringer Guertel 18-20, 1090 Vienna, Austria.

Published: June 2011

Background: In several clinical situations, lung separation and single-lung ventilation (SLV) is essential. In these cases, the double-lumen tube (DLT) is the most widely used device. Bronchial blocker such as Univent or Arndt Blocker serves as an alternative. The EZ-Blocker(®) (EZ; AnaesthetIQ B.V., Rotterdam, The Netherlands) is a new device promising to exceed clinical performance of DLT. The aim of this study was to assess the clinical performance of EZ in comparison with conventional left-sided DLT.

Methods: Forty adult patients undergoing elective thoracic surgery requiring thoracotomy and SLV were included in this study. The patients were randomly assigned to one of two groups: EZ (combined with conventional 7.5 or 8.5 mm single-lumen tube) or DLT (37 or 39 Fr left-sided DLT). Time for intubation procedure and time to verification of the correct position of EZ or DLT using fibreoptic bronchoscopy (FOB) were recorded. After surgery, a thoracic surgeon rated the quality of collapse of the lung (1-3 on a three-level scale).

Results: Time for intubation using DLT 85.5 (54.8) s was significantly faster (P<0.001) than using EZ 192 (89.7) s, whereas time for bronchoscopy was not significantly different (P=0.556). Conditions of surgery were rated equally [DLT 1.3 (0.6) vs EZ 1.4 (0.6), P=0.681].

Conclusions: Although time for intubation was longer with the EZ, the device proved to be an efficient and easy-to-use device. The EZ is a valuable alternative device to conventional DLT. Verification of the correct position of the EZ by FOB seems to be obligatory. This study was registered at http://www.clinicaltrials.gov (identifier: NCT01171560).

Download full-text PDF

Source
http://dx.doi.org/10.1093/bja/aer086DOI Listing

Publication Analysis

Top Keywords

double-lumen tube
8
single-lung ventilation
8
tube dlt
8
clinical performance
8
time intubation
8
dlt
6
randomized clinical
4
clinical trial
4
trial comparing
4
comparing double-lumen
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!