Purpose: Prostate carcinoma is a commonly diagnosed cancer in men. Nonsurgical treatment of early stage prostate cancer is an important alternative. The use of MRI for tumor cryoablation is of particular interest: it offers lower morbidity compared with other localized techniques. However, the current manual procedure is very time-consuming and has limited accuracy. A novel robotic assistant is therefore designed for prostate cancer cryotherapy treatment under MRI guidance to improve efficiency and accuracy.

Methods: Gesture definition was achieved based on actions of interventional radiologists at University Hospital of Strasbourg. A transperineal approach with a semiautonomous prostatic cryoprobe localization procedure was developed where the needle axis is automatically positioned before manual insertion. The workflow was developed simultaneously with the robotic assistant used for needle positioning.

Results: The design and the associated workflow of an original wire-driven manipulator were developed. The device is compact and has a low weight: its overall dimensions in the scanner are 100 × 100 × 40 mm with a weight of 120 g. Very good MRI compatibility was demonstrated.

Conclusions: A novel cryoablation procedure based on the use of a robotic assistant is proposed. The device design was presented with demonstration of MRI compatibility. Further developments include automatic registration and in vivo experimental testing.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s11548-011-0558-4DOI Listing

Publication Analysis

Top Keywords

robotic assistant
12
prostate cancer
8
mri compatibility
8
mri
5
design considerations
4
considerations novel
4
novel mri
4
mri compatible
4
compatible manipulator
4
prostate
4

Similar Publications

Background: To assess how centralisation of cancer services via robotic surgery influenced positive surgical margin (PSM) occurrence and its associated risk of biochemical recurrence (BCR) in cases of pT2 prostate cancer (PC).

Methods: Retrospective analysis of all radical prostatectomy (RP) cases performed in the West of Scotland during the period from January 2013 to June 2022. Primary outcomes were PSM and BCR.

View Article and Find Full Text PDF

Purpose: Cochlear implants (CI) are the most successful bioprosthesis in medicine probably due to the tonotopic anatomy of the auditory pathway and of course the brain plasticity. Correct placement of the CI arrays, respecting the inner ear anatomy are therefore important. The ideal trajectory to insert a cochlear implant array is defined by an entrance through the round window membrane and continues as long as possible parallel to the basal turn of the cochlea.

View Article and Find Full Text PDF

Robot-assisted vs. manual cochlear implant electrode array insertion in four children.

Eur Arch Otorhinolaryngol

January 2025

Department of Audiovestibology, ASST dei Sette Laghi, Via Lazio, 21100, Varese, VA, Italy.

Purpose: Evaluate the feasibility and safety of a robotic electrode insertion in pediatric cochlear implantation and compare the results with manually inserted electrodes in the same subject.

Methods: Retrospective case series review of four children who underwent bilateral cochlear implantation with the same array: on one side, the array was inserted using the robot, while on the other side the array was inserted manually. Behavioural and electrophysiological measures were compared.

View Article and Find Full Text PDF

Ankle push-off is important for efficient, human-like walking, and many prosthetic devices mimic push-off using motors or elastic elements. The knee is extended throughout the stance phase and begins to buckle just before push-off, with timing being crucial. However, the exact mechanisms behind this buckling are still unclear.

View Article and Find Full Text PDF

Safety and Feasibility of Robotic Nipple-Sparing Mastectomy With Immediate Direct-to-Implant Reconstruction - Insights From the One of the Largest Centers in Asia.

Clin Breast Cancer

December 2024

Department of General Surgery, Shin Kong Wu Ho-Su Memorial Hospital, Shilin, Taipei, Taiwan; Breast Cancer Center, Shin Kong Wu Ho-Su Memorial Hospital, Shilin, Taipei, Taiwan; School of Medicine, Fu Jen Catholic University, New Taipei City, Taiwan; School of Medicine, College of Medicine, Fu Jen Catholic University, Taipei, Taiwan. Electronic address:

Background: The use of robotic-assisted nipple-sparing mastectomy (R-NSM) with immediate direct-to-implant (DTI) reconstruction in treatment of breast cancer has been a controversial topic. The adoption of robotic surgery in breast cancer treatment has gained traction globally due to its minimally invasive nature, potential for improved cosmetic outcomes and better intraoperative visualization. This study provides insights on safety and feasibility robotic mastectomy at one of the largest centers in Asia.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!