Design of manikin for testing of residual-limb shape-capture method: technical note.

J Rehabil Res Dev

The National Centre for Prosthetics and Orthotics, University of Strathclyde, Glasgow, UK.

Published: September 2011

Consensus is still lacking on how best to capture the shape of a residual limb. Computer-aided design systems have not proven more accurate, repeatable, or reliable than traditional plaster of paris methods. Research is limited in design, relates to clinical trials, and is based on opinions and clinical experience. Many outcome measurements are based on qualitative estimations of prosthetic fit or patient feedback rather than quantitative measurements. Research must identify the most accurate, repeatable, and reliable methods for residual-limb shape capture under conditions most likely to enhance socket fit. Measurement is difficult because a reference grid is required for identifying the residual limb's axis for ensuring direct comparison. This article describes a manikin production method for testing the shape capture of the residual limb. Diameters and volume were measured at specific levels with a programmable computer numerical control milling machine and a displacement tool, with a combined accuracy of 5 micrometers.

Download full-text PDF

Source
http://dx.doi.org/10.1682/jrrd.2010.06.0113DOI Listing

Publication Analysis

Top Keywords

residual limb
8
accurate repeatable
8
repeatable reliable
8
shape capture
8
design manikin
4
manikin testing
4
testing residual-limb
4
residual-limb shape-capture
4
shape-capture method
4
method technical
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!