The performance of broadband volume holographic imaging system in terms of depth selectivity is investigated. The mechanism for depth resolution degradation is explained. In order to overcome this resolution degradation, a novel imaging device, the confocal-rainbow volume holographic imaging system, is proposed. Modeling and experimental validation of the performance of this novel imaging system indicates that depth resolution <16 μm is achievable. The lateral resolution of this device is <2.5 μm along a field of view of 300 μm×100 μm.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3076630 | PMC |
http://dx.doi.org/10.1364/AO.50.001382 | DOI Listing |