Small diameter transosseous wires are main parts of the Ilizarov frame concept. Wire tensioning is essential to gain the necessary transversal stiffness, and the wire fixators are therefore important, coupling the wire to the ring. The ability of three different wire fixator designs to maintain wire tension under dynamic loading is described. The three fixator designs were combinations of two types of bolts and one washer. In a simplified frame set-up consisting of one ring and one wire, sixty wires were dynamically loaded in sequence with 200N for 450 times and wire tension was repeatedly measured. The tension loss employing the different fixator types at two wire pretension levels (883N, 1275N) was measured. Wire slippage in the fixators and the midpoint deflection angle of the wires were measured. Large differences in fixation ability were observed. Wire fixators holding a larger and rougher contact surface had significant smaller tension loss compared to fixators with a smaller and smother contact surface. Both plastic wire deformation and slippage causes tension loss with slippage as the main contributor. Wire fixators' design can be improved in order to maintain wire tension. Such improvement could be of clinical interest enhancing the mechanical stability of the fixator.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jbiomech.2011.03.018 | DOI Listing |
Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi
January 2025
Department of Orthopedics, the First Affiliated Hospital of Bengbu Medical University, Bengbu Anhui, 233000, P. R. China.
Objective: To analyze the effectiveness of three internal fixation methods, namely hollow screw combined with Kirschner wire tension band, hollow screw combined with anchor nail, and modified 1/3 tubular steel plate, in the treatment of avulsion fracture of tibial tubercle (AFTT) in adolescents.
Methods: Between January 2018 and September 2023, 19 adolescent AFTT patients who met the selection criteria were admitted. According to different internal fixation methods, patients were divided into group A (8 cases, hollow screw combined with Kirschner wire tension band), group B (6 cases, hollow screw combined with anchor nail), and group C (5 cases, modified 1/3 tubular steel plate).
Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi
January 2025
Department of Joint and Bone Surgery, the First Affiliated Hospital of Wannan Medical College, Wuhu Anhui, 241000, P. R. China.
Objective: To explore the feasibility and effectiveness of suture anchor double-pulley technique combined with suture three-dimensional binding via bone tunnel technique for avulsion fractures of the inferior pole of the patella.
Methods: A clinical data of 38 patients with avulsion fractures of the inferior pole of the patella, who met the selective criteria and were admitted between September 2021 and April 2023, was retrospectively analyzed. The fractures were treated with suture anchor double-pulley technique combined with suture three-dimensional binding via bone tunnel technique in 18 cases (group A) and steel wire tension-band fixation in 20 cases (group B).
Orthop Surg
January 2025
Department of Orthopedics Surgery, The Second Affiliated Hospital of Soochow University, Suzhou, Jiangsu Province, China.
Objective: Inferior pole patellar fractures (IPPFs) pose a significant challenge due to their complex fracture patterns and high risk of complications associated with current treatment methods. This study aims to (1) characterize the fracture patterns of IPPFs using fracture mapping and (2) compare the biomechanical stability and clinical outcomes of treatment with anchor suture with patellar cerclage versus Kirschner-wire tension band combined with patellar cerclage.
Methods: (1) A retrospective analysis was conducted on 61 patients with IPPF.
JBJS Essent Surg Tech
January 2025
Department of Neurosurgery, Center for Neuroscience and Spine, Virginia Mason Medical Center, Seattle, Washington.
Background: Prone transpsoas lumbar interbody fusion (PTP) is a newer technique to treat various spinal disc pathologies. PTP is a variation of lateral lumbar interbody fusion (LLIF) that is performed with the patient prone rather than in the lateral decubitus position. This approach offers similar benefits of lateral spinal surgery, which include less blood loss, shorter hospital stay, and quicker recovery compared with traditional open spine surgery.
View Article and Find Full Text PDFBMC Musculoskelet Disord
January 2025
Department of Orthopedic Surgery, The Second Hospital Affiliated to Shenyang Medical College, Shenyang City, Liaoning Province, China.
Objective: The treatment of comminuted inferior pole patellar fractures has long posed a challenge for orthopedic surgeons. This study aims to compare the biomechanical stability and clinical efficacy of Kirschner wire tension band combined with anchor cross-suture fixation versus traditional partial patellectomy in the treatment of comminuted inferior pole patellar fractures.
Methods: A retrospective analysis was conducted on 14 patients who underwent Kirschner wire tension band combined with anchor cross-suture fixation (Group A) in our department of orthopedics from September 2020 to April 2022.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!