Performance of new enzyme-linked fluorescent assays for detection of Epstein-Barr virus specific antibodies in routine diagnostics.

Wien Klin Wochenschr

Department of Virology and Serological Diagnostics, Institute of Hygiene, Microbiology and Environmental Medicine, Medical University of Graz, Austria.

Published: April 2011

Background And Aims: Indirect fluorescent antibody assays are still the gold standard for the detection of Epstein-Barr-virus-specific antibodies; however, this technique requires several manual steps and an experienced technician. This retrospective study investigated the performance of the new VIDAS(®) enzyme-linked fluorescent assays for automated qualitative detection of EBV VCA IgM, VCA/EA IgG, and EBNA IgG antibodies in routine diagnostics.

Methods: 174 serum samples were tested first with the gold standard. In context with the clinical status, 60 samples IFA IgG/IgM positive and with clinical symptoms compatible with infectious mononucleosis, 26 samples IFA IgG/IgM negative and missing any clinical symptoms and 88 samples with varying IFA status and without any clinical information were defined. In a second step all samples were retested with the new assays.

Results: In the overall agreement between VIDAS(®) and IFA for evaluable results, almost perfect agreement was observed (kappa = 0.91; 95% confidence interval (CI), 0.86-0.97). Estimating all indeterminate VIDAS(®) results as discordant or concordant the observed kappa values were 0.71 (CI 0.63-0.79) and 0.93 (CI 0.88-0.98), respectively. With the new assays 45, 22, and 70 identical results were obtained, respectively. Western blot analysis of the discrepant samples showed a quasi similar performance of both assays.

Conclusions: The new VIDAS(®) assays can be an alternative to IFA testing especially in high-throughput laboratories. Full automation of EBV serological diagnostis by the new VIDAS assays is of major importance for routine diagnostic laboratories.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00508-011-1561-zDOI Listing

Publication Analysis

Top Keywords

enzyme-linked fluorescent
8
fluorescent assays
8
antibodies routine
8
gold standard
8
samples ifa
8
ifa igg/igm
8
clinical symptoms
8
observed kappa
8
assays
6
samples
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!