Introduction: To determine six-degree of freedom of total knee arthroplasty kinematics (TKA), optimized matching algorithms for single fluoroscopic image system may be used. Theoretical accuracy of these systems was reported. Nevertheless, all reports were done under idealized laboratory experimental conditions. The aim of this study was to evaluate the "true" accuracy of a flat panel single plane video-fluoroscopy system based on computed-assisted design (CAD) model matching and compare it to TKA kinematics obtained from optoelectronic measurements as gold standard.

Hypothesis: The estimation of the error produced by 2D/3D fluoroscopic registration in daily practice is misjudged in most available laboratory reports.

Material And Methods: The experimental set-up used a TKA implanted into femoral and tibial cadaver bones. Thirty flexions were simultaneously registered using single plane fluoroscopy and an active optical tracking system. Kinematics registered were compared using the root mean square error (RMS), the concordance correlation coefficient and Bland & Altman plot analysis.

Results: The mean range of motion of flexion during the experiment was 106°. The respective RMS for flexion, varus-valgus and internal-external rotation were 0.68, 0.67 and 1.02°. The respective RMS for antero-posterior, medio-lateral and proximo-distal displacement were 1.3, 2.4 and 1.06 mm. Extreme values of the measured error concerning medio-lateral displacement were -5.4 and 22,1mm.

Discussions: Analysis found some outliners in all degree of freedom with a systematic error and larger standard deviation than already published data. One should make sure that during the experiment the motion of interest is in the in-plane direction. Moreover, this study brings out the true threshold detection of this type of analysis.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.otsr.2011.01.003DOI Listing

Publication Analysis

Top Keywords

total knee
8
knee arthroplasty
8
single plane
8
respective rms
8
arthroplasty three-dimensional
4
three-dimensional kinematic
4
kinematic estimation
4
estimation prevision
4
prevision two-dimensional
4
two-dimensional fluoroscopy
4

Similar Publications

Purpose: To evaluate the radiological and clinical outcomes in two patient groups: first, varus aligned medial meniscus posterior root tear (MMPRT) patients who underwent posteromedial open wedge high tibial osteotomy (PMOWHTO) and simultaneous root repair; second, patients with varus medial knee osteoarthritis without MMPRT who underwent PMOWHTO.

Methods: Patients had MMPRT repair concomitant with PMOWHTO and varus medial knee osteoarthritis without concomitant root tear patients who underwent PMOWHTO and were reviewed. Radiographic parameters, medial meniscus extrusion (MME) and Knee Society Scores [KSSs, including the following subscores: knee score (KS) and knee function score (KFS)] were evaluated.

View Article and Find Full Text PDF

Background: This study investigated the influence of surgical alignment techniques on knee joint biomechanics during stair negotiation tasks. Our hypothesis was that a more personalized joint alignment would result in reduced medial knee loading biomechanics to negotiate the stairs.

Methods: There were 28 adults (14 mechanical alignments [MA], 14 kinematic alignment [KA]) who underwent total knee arthroplasty (TKA) at least one year post-operatively and performed five stair ascent and descent trials at their preferred velocities.

View Article and Find Full Text PDF

Introduction: As the number of revision total knee arthroplasties (rTKA) continues to rise, there is increasing interest in the use of contemporary rotating hinge prostheses. These devices often incorporate porous cones to fill bone defects and enhance long-term fixation. This study evaluated the clinical and functional outcomes and survivorship in rTKA patients utilizing a rotating hinge prosthesis with flexible titanium (FT) cones, porous tantalum (PT) cones, or no cones.

View Article and Find Full Text PDF

Introduction: Flexion instability (FI) is increasingly being recognized as a common reason for failure after total knee arthroplasty (TKA). However, the diagnosis remains loosely defined and is often described by a constellation of symptoms and findings with the exclusion of other clinical entities. The purpose of this study was to examine the definition and diagnostic criteria of FI used in the current peer-reviewed literature.

View Article and Find Full Text PDF

Introduction: Knee alignment significantly impacts the outcome of total knee arthroplasty (TKA). Understanding patient perceptions of their knee alignment in relation to objective measurements is essential to ensure optimal surgical outcomes and to meet patients' expectations. This study reports patients' perception of pre- and postoperative knee alignment in relation to radiographic alignment measurements.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!