A method was developed to calculate the second dimension retention index of comprehensive two-dimensional gas chromatography time-of-flight mass spectrometry (GC×GC/TOF-MS) data using n-alkanes as reference compounds. The retention times of the C(7)-C(31) alkanes acquired during 24 isothermal experiments cover the 0-6s retention time area in the second dimension retention time space, which makes it possible to calculate the retention indices of target compounds from the corresponding retention time values without the extension of the retention space of the reference compounds. An empirical function was proposed to show the relationship among the second dimension retention time, the temperature of the second dimension column, and the carbon number of the n-alkanes. The proposed function is able to extend the second dimension retention time beyond the reference n-alkanes by increasing the carbon number. The extension of carbon numbers in reference n-alkanes up to two more carbon atoms introduces <10 retention index units (iu) of deviation. The effectiveness of using the proposed method was demonstrated by analyzing a mixture of compound standards in temperature programmed experiments using 6 different initial column temperatures. The standard deviation of the calculated retention index values of the compound standards fluctuated from 1 to 12 iu with a mean standard deviation of 5 iu.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3078172PMC
http://dx.doi.org/10.1016/j.chroma.2011.02.072DOI Listing

Publication Analysis

Top Keywords

second dimension
24
dimension retention
20
retention time
20
retention
10
retention comprehensive
8
comprehensive two-dimensional
8
two-dimensional gas
8
gas chromatography
8
chromatography time-of-flight
8
time-of-flight mass
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!