Rationale And Objectives: The clinimetric properties of blood oxygen level‒dependent (BOLD) magnetic resonance imaging (MRI) for assessment of musculoskeletal changes have been poorly investigated. The study objectives were to assess the interframework reliability of data acquisition of BOLD MRI and to test its convergent validity in chronic arthritis in a rabbit model of inflammatory arthritis as compared with corresponding clinical and laboratory measures.
Materials And Methods: One of the knees of 12 New Zealand male white rabbits was injected with a 1% carrageenin solution, and the contralateral (control) one was not. Twelve rabbits were euthanized on day 28 of arthritis (chronic arthritis). Clinical (joint diameters), laboratory (serum amyloid A concentration), and BOLD MRI measurements were obtained on days 0, 1, and 28 of arthritis. Twenty paradigms of data acquisition and analysis were applied.
Results: The most reliable MRI parameters set, regardless of threshold values used for data analysis, was spiral technique (level 1), 40 ms of echo time (level 2), 60 seconds of on_ and off_ paradigm (level 3) and carbogen mixture of gases (95% O2 + 5% CO2) (level 4). With regard to construct validity, BOLD imaging correlated moderately (r = -.54, P < .0001) with knee diameters, and weakly (r = -.35, P = .01) with laboratory indices (high threshold for analysis).
Conclusion: BOLD MRI has a substantial or excellent interframework reliability for assessment of arthritic rabbit knees; however, it correlates only moderately or poorly with clinical and laboratory measures. Nevertheless, this study supports further validation of BOLD MRI for assessment of soft tissue changes in a rabbit model of arthritis.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.acra.2010.12.008 | DOI Listing |
Sci Rep
January 2025
Department of Radiology, Medical College of Wisconsin, 8701 Watertown Plank Road, Milwaukee, WI, 53226, USA.
The alteration of neurovascular coupling (NVC), where acute localized blood flow increases following neural activity, plays a key role in several neurovascular processes including aging and neurodegeneration. While not equivalent to NVC, the coupling between simultaneously measured cerebral blood flow (CBF) with arterial spin labeling (ASL) and blood oxygenation dependent (BOLD) signals, can also be affected. Moreover, the acquisition of BOLD data allows the assessment of resting state (RS) fMRI metrics.
View Article and Find Full Text PDFRofo
January 2025
Department of Diagnostic and Interventional Radiology, University Hospital Tübingen, Tübingen, Germany.
Multiparametric MRI is a promising technique for noninvasive structural and functional imaging of the kidneys that is gaining increasing importance in clinical research. Still, there are no standardized recommendations for analyzing the acquired images and there is a need to further evaluate the accuracy and repeatability of currently recommended MRI parameters. The aim of the study was to evaluate the test-retest repeatability of functional renal MRI parameters using different image analysis strategies.
View Article and Find Full Text PDFAnesthesiology
January 2025
Department of Neuroradiology, School of Medicine and Health, Technical University of Munich, Germany.
Background: According to the model of the glymphatic system, the directed flow of cerebrospinal fluid (CSF) is a driver of waste clearance from the brain. In sleep, glymphatic transport is enhanced, but it is unclear how it is affected by anesthesia. Animal research indicates partially opposing effects of distinct anesthetics but corresponding results in humans are lacking.
View Article and Find Full Text PDFElife
January 2025
Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Harvard Medical School, Massachusetts General Hospital, Charlestown, United States.
High-resolution awake mouse functional magnetic resonance imaging (fMRI) remains challenging despite extensive efforts to address motion-induced artifacts and stress. This study introduces an implantable radio frequency (RF) surface coil design that minimizes image distortion caused by the air/tissue interface of mouse brains while simultaneously serving as a headpost for fixation during scanning. Furthermore, this study provides a thorough acclimation method used to accustom animals to the MRI environment minimizing motion-induced artifacts.
View Article and Find Full Text PDFCogn Neurodyn
December 2025
Department of Radiology, The 960th Hospital of People's Liberation Army Joint Logistic Support Force, Jinan, China.
Insomnia is a common mental illness seriously affecting people lives, that might progress to major depression. However, the neural mechanism of patients with CID comorbid MDD remain unclear. Combining fractional amplitude of low-frequency fluctuation (fALFF) and seed-based functional connectivity (FC), this study investigated abnormality in local and long-range neural activity of patients with CID comorbid MDD.