We experienced a case with a falsely low value by a blood MMP-3 measuring reagent employing a recently structured new latex immunoturbidity. The case involved duplicate orders for one patient in a single day, and the blood collection amounts and measured values were approximately 6.0 mL and 206.6 ng/mL and approximately 1.0 mL and 107.5 ng/mL. The latter MMP-3 concentration was 48% of the former, showing a low tendency. Therefore, an experiment was conducted by adding serum to the blood collection tubes with or without a serum-separating agent of four different manufacturers (Terumo, Sekisui Medical, Nipro, and Becton Dickinson), and similar results as our experienced case were obtained with the Terumo tube with serum-separating agent, which had been used in this case. The amount of whole blood was obtained by conversion assuming a hematocrit value of 40%, and the addition ratio was calculated relative to the predetermined amount of the tube showing a falsely low value. Falsely low values were observed at < or = 56%, < or = 21%, < or = 20%, and , or = 33% for Terumo, Sekisui Medical, Nipro, and Becton Dickinson, with tubes containing a serum-separating agent, and at < or =10%, < or =8%, < or =19%, and < or =14% for Terumo, Sekisui Medical, Nipro, and Becton Dickinson, respectively, with plain tubes. Falsely low values were not observed with the 10-ml plain tube of Terumo and the 9-ml plain tube of Nipro (untreated tube). Based on these results, care should be taken if samples are below the predetermined amount of the blood collection tube to determine the serum MMP-3 by this method.

Download full-text PDF

Source

Publication Analysis

Top Keywords

falsely low
16
blood collection
12
serum-separating agent
12
terumo sekisui
12
sekisui medical
12
medical nipro
12
nipro becton
12
becton dickinson
12
experienced case
8
tubes serum-separating
8

Similar Publications

Fit for purpose testing and independent GMP validation of the monocyte activation test.

Curr Res Toxicol

December 2024

Large Molecule Analytical Development, Janssen Biologics LLC (a Johnson and Johnson Company), 200 Great Valley Parkway, Malvern, PA 19355 USA.

The present study describes the "fit for purpose" testing and the independent product-specific GMP validation of the monocyte activation test (MAT) to detect pyrogenic and pro-inflammatory contaminants, MAT Method A, Quantitative Test (European Pharmacopoeia, Ph. Eur. chapter 2.

View Article and Find Full Text PDF

Purpose Owing to the shortage of surgeons and the decrease in medical staff in regional medical care, reducing unnecessary tests can limit the burden on the staff. In this study, we aimed to examine the predictors of deep vein thrombosis (DVT), such as D-dimer levels in patients who underwent surgery at our hospital, and determine the feasibility of screening in these patients. Knowledge of D-dimer levels can indicate the risk of DVT in patients about to undergo surgery.

View Article and Find Full Text PDF

Performing hypothesis tests with adequate statistical power is indispensable for psychological research. In response to several large-scale replication projects following the replication crisis, concerns about the root causes of this crisis - such as questionable research practices (QRPs) - have grown. While initial efforts primarily addressed the inflation of the type I error rate of research due to QRPs, recent attention has shifted to the adverse consequences of low statistical power.

View Article and Find Full Text PDF

Background: Sickle cell disease (SCD) confers neurological risks that contribute to cognitive and academic difficulties. Clinical guidelines state that cognition should be monitored using signaling questions. However, evidence is lacking regarding the extent to which signaling questions accurately identify children with cognitive issues.

View Article and Find Full Text PDF

Lysosome-targeted dual-locked NIR fluorescent probe for visualization of HS and viscosity in drug-induced liver injury and tumor models.

Anal Chim Acta

February 2025

Department of Radiology, Changhai Hospital, Naval Medical University, Changhai Road 168, Shanghai 200433, China. Electronic address:

Background: Lysosomes, as an indispensable subcellular organelle have numerous physiological functions closely associated with HS and viscosity, and accurate assessment of HS/viscosity fluctuations in lysosomes is essential for gaining a comprehensive understanding of lysosome-related physiological activities and pathological processes. The previous single-response fluorescent probes for either HS or viscosity alone have the potential to generate "false positive" signals in a complex biological environment. In contrast, dual-locked probes can simultaneously respond to multiple targets simultaneously, which could effectively eliminate this defect.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!