Non-invasive measurements of cardiac output in atrial fibrillation: inert gas rebreathing and impedance cardiography.

Scand J Clin Lab Invest

Department of Cardiology, Hvidovre Hospital, Faculty of Health Sciences, University of Copenhagen, Copenhagen, Denmark.

Published: July 2011

AI Article Synopsis

Article Abstract

Background: Atrial fibrillation (AF) is associated with significant morbidity and mortality. To test the effect of interventions, knowledge of cardiac output (CO) is important. However, the irregular heart rate might cause some methods for determination of CO to have inherent weaknesses. Objective. To assess the validity of these methods in AF, a new inert gas rebreathing device and impedance cardiography was tested with echocardiography as reference.

Methods: Using a cross-sectional design, 127 patients with AF and 24 in SR were consecutively recruited. Resting CO was measured using inert gas rebreathing (n = 62) or impedance measurement of intrathoracic blood flow (n = 89) in separate studies with echocardiographic measurement as reference.

Results: CO determined with impedance cardiography was mean 4.77 L/min ± 2.24(SD) compared to 4.93 L/min ± 1.17 by echocardiography (n = 89, n.s.) in patients with AF. CO by inert gas rebreathing was 4.98 L/min ± 2.49(SD) compared to 5.70 L/min ± 2.49 by echocardiography (n = 62, n.s.) in patients with AF and SR (AF 5.42 ± 2.9 vs. 6.27, n.s. and SR 4.09 ± 1.08 vs. 4.35 ± 0.86, n.s.). Mean bias between impedance cardiography and echocardiography was 0.14 ± 0.95 L/min and -0.13 ± 0.98 L/min between inert gas rebreathing and echocardiography. Inert gas rebreathing showed larger intra-patient variation than impedance cardiography (0.11 vs. 0.054). Correlation between inert gas rebreathing and echocardiography was r = -0.060 and between impedance cardiography and echocardiography was r = 0.128. Impedance cardiography and inert gas rebreathing both underestimated CO compared to echocardiography.

Conclusion: Variation between the inert gas rebreathing and the reference method for AF patients was less than desired. Impedance cardiography was superior to inert gas rebreathing and showed acceptable agreement with echocardiography and variability similar to echocardiography.

Download full-text PDF

Source
http://dx.doi.org/10.3109/00365513.2011.563790DOI Listing

Publication Analysis

Top Keywords

inert gas
40
gas rebreathing
40
impedance cardiography
32
inert
10
gas
10
rebreathing
10
impedance
9
echocardiography
9
cardiac output
8
atrial fibrillation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!