[Analysis on DQA protocol of fMRI].

Sheng Wu Yi Xue Gong Cheng Xue Za Zhi

Department of Radiology of West China Hospital, Sichuan University, Chengdu 610041, China.

Published: December 2010

Our purpose is to introduce and analyze the data quality assurance (DQA) protocol of functional magnetic resonance imaging (fMRI). A water phantom was scanned to get DQA indexes. An fMRI sequence was used to get signal noise ratio (SNR) and Drift, which was calculated from maximum difference ratio of the average signal intensity in the region of interest (ROI) of image serials. The long period application of this method demonstrated that this DQA protocol can reflect imaging performance and the state of stability of the MRI scanner. Some application experience and discussion involved in DQA were also presented here.