A liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed to determine total N-acetylcysteine in human plasma. Mass spectrometric detection was achieved in positive electrospray ionization and multiple reaction monitoring mode. The mass transition pairs of N-acetylcysteine and the isotope-labeled internal standard d3-N-acetylcysteine were 164 → 122 and 167 → 123, respectively. The method was linear over the range of 10-5000 ng/mL in human plasma. The adoption of trichloroacetic acid significantly enhanced the extraction recovery. The blank matrix was screened to minimize the influence of endogenous N-acetylcysteine. After being fully validated, the method was successfully applied to the pharmacokinetic and bioequivalent study of N-acetylcysteine after oral administration of 600 mg tablets to 24 healthy Chinese volunteers.

Download full-text PDF

Source
http://dx.doi.org/10.1002/bmc.1465DOI Listing

Publication Analysis

Top Keywords

human plasma
12
mass spectrometry
8
n-acetylcysteine human
8
isotope-labeled internal
8
internal standard
8
n-acetylcysteine
5
liquid chromatography
4
chromatography tandem
4
mass
4
tandem mass
4