Background: The insertion of Ventricular Assist Devices is a common strategy for cardiovascular support in patients with refractory cardiogenic shock. This study sought to determine the impact of ventricular assist devices on the dynamic relationship between arterial blood pressure and cerebral blood flow velocity.
Methods: A sample of 5 patients supported with a pulsatile ventricular assist device was compared with 5 control patients. Controls were matched for age, co-morbidities, current diagnosis and cardiac output state, to cases. Beat-to-beat recordings of mean arterial pressure and cerebral blood flow velocity, using transcranial Doppler were obtained. Transfer function analysis was performed on the lowpass filtered pressure and flow signals, to assess gain, phase and coherence of the relationship between mean arterial blood pressure and cerebral blood flow velocity. These parameters were derived from the very low frequency (0.02-0.07 Hz), low frequency (0.07-0.2 Hz) and high frequency (0.2-0.35 Hz).
Results: No significant difference was found in gain and phase values between the two groups, but the low frequency coherence was significantly higher in cases compared with controls (mean ± SD: 0.65 ± 0.16 vs 0.38 ± 0.19, P = 0.04). The two cases with highest coherence (~0.8) also had much higher spectral power in mean arterial blood pressure.
Conclusions: Pulsatile ventricular assist devices affect the coherence but not the gain or phase of the cerebral pressure-flow relationship in the low frequency range; thus whether there was any significant disruption of cerebral autoregulation mechanism was not exactly clear. The augmentation of input pressure fluctuations might contribute in part to the higher coherence observed.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3052186 | PMC |
http://dx.doi.org/10.1186/1471-2253-11-4 | DOI Listing |
Front Cardiovasc Med
December 2024
Department of Cardiovascular Surgery, Xijing Hospital, Air Force Medical University, Xi'an, Shaanxi, China.
Background: To evaluate the feasibility, effectiveness and assistant effect of 3D printed aortic model in the treatment on congenital coarctation of the aorta (CoA) in adolescents and adults.
Methods: From December 2018 to December 2023, a total of 10 patients with congenital coarctation of aorta underwent percutaneous balloon dilatation covered stent implantation in the department of cardiovascular surgery, Xijing Hospital. There were 6 males and 4 females whose average age was (27.
J Neurosurg
December 2024
2Nell Hodgson Woodruff School of Nursing, Emory University, Atlanta; and.
Objective: The objective was to evaluate the etiology, natural history, and impact of surgical intervention on outcomes of left ventricular assist device (LVAD) patients presenting with intracranial hemorrhage (ICH).
Methods: The authors completed a retrospective review of LVAD patients who presented with ICH at 2 centers between 2013 and 2022. Patients were reviewed for demographic, clinical, and radiographic variables.
PLoS One
December 2024
Department of Radiology, Columbia University, New York, NY, United States of America.
Background: Imaging both electrical and mechanical cardiac function can better characterize cardiac disease and improve patient care. Currently, there is no noninvasive technique that can simultaneously image both electrical and mechanical function of the whole heart at the point of care. Here, our aim is to demonstrate that high volume-rate echocardiography can simultaneously map cardiac electromechanical activation and end-systolic cardiac strain of the whole heart in a single heartbeat.
View Article and Find Full Text PDFJ Clin Monit Comput
December 2024
Department of Anesthesia and Intensive Care, "Policlinico San Marco" University Hospital, Catania, Italy.
Echocardiography is crucial for evaluating patients at risk of clinical deterioration. Left ventricular ejection fraction (LVEF) and velocity time integral (VTI) aid in diagnosing shock, but bedside calculations can be time-consuming and prone to variability. Artificial intelligence technology shows promise in providing assistance to clinicians performing point-of-care echocardiography.
View Article and Find Full Text PDFJ Cardiovasc Dev Dis
December 2024
3rd Department of Cardiology, "Sotiria" Chest Diseases Hospital, Medical School, National and Kapodistrian University of Athens, 11527 Athens, Greece.
Hypertrophic cardiomyopathy (HCM) is a complex and heterogeneous cardiac disorder, often complicated by cardiogenic shock, a life-threatening condition marked by severe cardiac output failure. Managing cardiogenic shock in HCM patients presents unique challenges due to the distinct pathophysiology of the disease, which includes dynamic left ventricular outflow tract obstruction, diastolic dysfunction, and myocardial ischemia. This review discusses current and emerging therapeutic strategies tailored to address the complexities of HCM-associated cardiogenic shock and other diseases with similar pathophysiology that provoke left ventricular outflow tract obstruction.