Practical device for precise cutting of costal cartilage grafts to uniform thickness.

Arch Facial Plast Surg

Beckman Laser Institute and Medical Clinic, University of California-Irvine, 1002 Health Sciences Road E, Irvine, CA 92612, USA.

Published: February 2012

Objectives: Costal cartilage is becoming increasingly popular as a graft source for facial reconstruction. However, carving methods have not changed in decades and continue to primarily rely on detailed maneuvers with a scalpel. There are few reports of mechanical devices for shaping costal cartilage, and to our knowledge their accuracy and precision have not been reported. We describe a simple costal cartilage slicing device that facilitates the production of sections having uniform, user-defined thicknesses.

Methods: The design included laboratory research using 200 porcine and 2 cadaveric human ex vivo costal cartilage slices. A 2-component apparatus was constructed consisting of a mechanism to secure the costal cartilage and a double-bladed device to cut the rib graft through a central cross-section. Optimizing blade characteristics and static forces that secure the cartilage were critical design challenges. The device was used to obtain slices 0.8, 2.1, and 4.1 mm in thickness, with lengths up to 4.0 cm and a width of 1.0 cm. To confirm uniformity, thickness was measured at 8 fixed regions per section using a digital micrometer.

Results: All costal cartilage slices appeared to be extremely uniform on visual and manual inspection. The absolute difference between the largest and smallest thickness measured for each individual sample ranged from 0.04 to 0.13 mm, 0.06 to 0.14 mm, and 0.10 to 0.21 mm for the 0.8-, 2.1-, and 4.1-mm-thick groups, respectively.

Conclusions: Our study demonstrates the precision of using a mechanical slicing device to section costal cartilage to a clinically relevant and uniform thickness. This mechanized technology may increase accuracy and reduce carving time required for using costal cartilage tissue in head and neck reconstruction.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4122103PMC
http://dx.doi.org/10.1001/archfacial.2011.8DOI Listing

Publication Analysis

Top Keywords

costal cartilage
36
cartilage
10
costal
9
uniform thickness
8
slicing device
8
cartilage slices
8
thickness measured
8
thickness
5
practical device
4
device precise
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!