Evaluation of new Forcefield technology: reducing platelet adhesion and cell coverage of pyrolytic carbon surfaces.

J Thorac Cardiovasc Surg

Division of Thoracic and Cardiovascular Surgery, Department of Surgery, University of Louisville, Louisville, KY, USA.

Published: October 2011

Introduction: Platelet adhesion and activation are a significant source of clinical complications. Preventing foreign surface-platelet interaction may improve biocompatibility of implantable medical devices. This study evaluated efficacy of novel technique for electrically modifying surface of conductive biomaterial and attaching blood components to prevent thrombogenesis. Specifically, this new surface modification technology, Forcefield (ATS Medical, Inc, Minneapolis, Minn), was designed to prevent platelet adhesion on pyrolytic carbon. A modulated low-voltage current is directly applied to pyrolytic carbon surfaces to stimulate adherence of a layer of charged proteins from circulating blood components that is resistive to platelet deposition.

Methods: Feasibility of Forcefield technology was tested in line with cardiopulmonary bypass circuit in patients undergoing standard cardiac surgery (n = 6). Forcefield treatment was applied to segment of pyrolytic carbon with 15 minutes (n = 3) and 30 minutes (n = 3) of electrically stimulated exposure time, and resulting segments were compared with untreated pyrolytic carbon segment. Platelet adhesion confluence was then quantified by scanning electron microscopy.

Results: Confluence of the Forcefield-treated pyrolytic carbon segments (3.3% ± 2.2%) was significantly reduced relative to untreated pyrolytic carbon control segments (81.7% ± 24%, P < .001). There were no discernible differences in cell confluence with Forcefield-treated segments as function of exposure time (15 or 30 minutes).

Conclusions: Forcefield technology may enable modification of pyrolytic carbon surfaces to prevent platelet adhesion and thrombogenesis of implanted medical devices, including heart valves, stents, catheters, and ventricular assist devices, and may eliminate the need for anticoagulant and antiplatelet therapies.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jtcvs.2011.01.012DOI Listing

Publication Analysis

Top Keywords

pyrolytic carbon
32
platelet adhesion
20
forcefield technology
12
carbon surfaces
12
pyrolytic
8
carbon
8
medical devices
8
blood components
8
prevent platelet
8
exposure time
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!