Computer navigation in total knee arthroplasty produces better component alignment than conventional techniques. Different navigation systems exist. We undertook a prospective, randomized study comparing 2 navigations systems (Stryker Full Navigation and Stryker Articular Surface Mounted [ASM] navigation systems). Three-dimensional component alignment (Perth computed tomographic knee protocol) and function at 1 year (Knee Society Scores) were assessed. Forty patients participated (20 fully navigated and 20 ASM-navigated total knee arthroplasties). Cohorts were well matched according to sex, age, and body mass index. No statistically significant difference was seen in any parameter of 3-dimensional component alignment or function between cohorts. Operative time for the ASM cohort was significantly less than the fully navigated cohort (P = .001). Both systems performed equally well, and therefore, surgeon preference should determine which system is used.