We describe the construction and performance of an NMR tube with a magnetic susceptibility matched sample cavity that confines the solution within the detection zone in the axial direction and in a quasi-rectangular region in the radial direction. The slot-like sample cavity provides both good sample volume efficiency and tolerance to sensitivity loss in the sample space. The signal-to-noise ratio per unit volume of the constructed tube was 2.2 times higher than that of a cylindrical tube of 5mm outer diameter with a sample containing 300 mM NaCl at a static magnetic field of 14.1T. Even the overall signal-to-noise ratio of the slot tube was 35% higher than that of the conventional 5mm tube for a sample containing 300 mM NaCl. Similar improvements over existing sample tube geometries were obtained at 950 MHz. Moreover the temperature rise resulting from RF heating was found to be significantly lower for the slot tube even when compared to 3 and 4mm outer diameter cylindrical tubes as measured in a 5mm cryoprobe. A further advantage of this type of tube is that a sample cavity of any desired size and shape can be formed within a cylindrical tube for use in a single cryogenic probe.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3065960PMC
http://dx.doi.org/10.1016/j.jmr.2011.01.005DOI Listing

Publication Analysis

Top Keywords

sample cavity
16
tube sample
12
tube
10
sample
9
construction performance
8
performance nmr
8
nmr tube
8
signal-to-noise ratio
8
cylindrical tube
8
outer diameter
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!