Purpose: To determine the accuracy of temporal artery and axillary temperatures and the discomfort level of stable neonates during temperature measurement.

Subjects: Convenience sample of neonates between the ages of 32 and 40 weeks' gestation cared for in an isolette or crib.

Design: A method-comparison design was used to compare different methods for noninvasive temperature monitoring (infra-red temporal artery; axillary electronic) to core body temperatures (indwelling rectal probe).

Main Outcome Measure: Bias and precision of test temperature devices (temporal artery; axillary).

Results: Bias and precision for the temporal artery and axillary devices were 0.30 ± 0.44 and 0.28 ± 0.33, respectively. Analysis of variance found significant differences between both temporal and axillary temperatures compared to rectal temperatures (P < .01). Statistical differences were small and did not represent a clinically important difference. No statistical difference was found between temporal artery and axillary temperatures (P = .81). Increases in neonate discomfort after temperature measurement were significantly greater with axillary than increases after temporal artery temperature measurement (P = .03).

Conclusions: This study found that body temperature measured with the temporal artery thermometer was similar to temperatures obtained with an axillary thermometer in stable, afebrile neonates. The use of temporal artery thermometry appears to be an acceptable approach for noninvasive temperature measurement in neonates, which causes less discomfort in neonates.

Download full-text PDF

Source
http://dx.doi.org/10.1097/ANC.0b013e3182087d2bDOI Listing

Publication Analysis

Top Keywords

temporal artery
36
artery axillary
16
axillary temperatures
12
temperature measurement
12
artery
9
temporal
9
accuracy temporal
8
artery thermometry
8
noninvasive temperature
8
bias precision
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!