Purpose: The influence of the numbers of projection of SPECT exerting on a re-constructed image cannot be strictly evaluated by phantom studies. Therefore, we compared re-constructed images of the FBP method and the ML-EM method by using simulation data.

Methods: Simulation data was entered in the image processing software, and the projection data that changed the numbers of projection was made. Afterwards, reconstructed images of the FBP and the ML-EM methods were compared with respect to contrast, %COV, and the NMSE value.

Result: When the numbers of projection of the FBP and the ML-EM method were decreased, all of the contrast, %COV, and the NMSE value were more deteriorated than that of the ideal image. Therefore, the image quality of SPECT improves with both FBP and ML-EM methods when there are many numbers of projection. Moreover, the FBP method was excellent in a cold contrast, and the ML-EM method was uniformly excellent. Therefore, an understanding of features and their inspection are effective for the selection of each image reconstruction method.

Download full-text PDF

Source
http://dx.doi.org/10.6009/jjrt.66.1587DOI Listing

Publication Analysis

Top Keywords

fbp ml-em
16
numbers projection
16
ml-em method
12
simulation data
8
images fbp
8
fbp method
8
ml-em methods
8
contrast %cov
8
%cov nmse
8
projection fbp
8

Similar Publications

Compressed sensing reconstruction shortens the acquisition time for myocardial perfusion imaging: a simulation study.

Radiol Phys Technol

September 2023

Department of Medical Quantum Science, Faculty of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka-shi, Fukuoka, 812-8582, Japan.

Article Synopsis
  • Compressed sensing (CS) has been explored for enhancing image quality in SPECT imaging, specifically for myocardial perfusion imaging (MPI), an area not extensively studied before.
  • This study compared CS-iterative reconstruction (CS-IR) against traditional methods: filtered back-projection (FBP) and maximum likelihood expectation maximization (ML-EM) in terms of acquisition time reduction and image quality.
  • The results showed that CS-IR achieved lower coefficients of variation (CV) for image uniformity and had the potential to significantly decrease acquisition time while maintaining similar contrast levels compared to ML-EM and FBP.
View Article and Find Full Text PDF
Article Synopsis
  • The study aimed to validate a new method for improving SPECT imaging using compressed sensing (CS) with offset acquisition to enhance image quality and accuracy.
  • Various reconstruction techniques were compared, showing that CS-IR with offset acquisition maintained high image quality even with fewer projections, while traditional methods like ML-EM suffered in performance.
  • The findings suggest that CS-IR with offset acquisition could lead to quicker SPECT scans without sacrificing image quality, benefiting both patients and imaging accuracy.
View Article and Find Full Text PDF

Hyperspectral X-ray CT (HXCT) technology provides not only structural imaging but also the information of material components therein. The main purpose of this study is to investigate the effect of various reconstruction algorithms on reconstructed X-ray absorption spectra (XAS) of components shown in the CT image by means of HXCT. In this paper, taking 3D printing polymer as an example, seven kinds of commonly used polymers such as thermoplastic elastomer (TPE), carbon fiber reinforced polyamide (PA-CF), acrylonitrile butadiene styrene (ABS), polylactic acid (PLA), ultraviolet photosensitive resin (UV9400), polyethylene terephthalate glycol (PETG), and polyvinyl alcohol (PVA) were selected as samples for hyperspectral CT reconstruction experiments.

View Article and Find Full Text PDF

Background: This work proposed a joint L1 and total variation (TV) regularized reconstruction method for X-ray fluorescence tomography (XFT), and investigated the performance of this method in quantitative imaging of gold nanoparticles (GNPs).

Methods: We developed a dual-modality XFT/CT imaging system which consisted of a benchtop X-ray source, a translation/rotation stage, a silicon drift detector for X-ray fluorescence (XRF) detection, and a flat panel detector for transmission X-ray detection. A pencil-beam collimator was 3D printed with steel and employed in sample excitation.

View Article and Find Full Text PDF

Impact of statistical reconstruction and compressed sensing algorithms on projection data elimination during X-ray CT image reconstruction.

Oral Radiol

September 2018

Department of Engineering on Intelligent Machines and Biomechanics, School of Regional Innovation and Social Design Engineering, Faculty of Engineering, Kitami Institute of Technology, 165 Koencho, Kitami, Hokkaido, 090-8507, Japan.

Objectives: To examine the effect of incomplete, or total elimination of, projection data on computed tomography (CT) images subjected to statistical reconstruction and/or compressed sensing algorithms.

Methods: Multidetector row CT images were used. The algebraic reconstruction technique (ART) and the maximum likelihood-expectation maximization (ML-EM) method were compared with filtered back-projection (FBP).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!