Objective: To describe the difficulties and differing techniques in the transcatheter placement of amplatz ventricular septal defect devices to close perimembranous ventricular septal defects and place these in the context of the expanding literature on ventricular septal defect catheter closure.

Background: Surgery remains the established first-line therapy for closure of haemodynamically significant perimembranous ventricular septal defects. Transcatheter techniques appeared to promise a possible alternative, obviating the need for cardiac surgery. However, significant technical and anatomical constraints coupled with ongoing reports of a high incidence of heart block have prevented these hopes from being realised to any significant extent. It is likely that there are important methodological reasons for the high complication rates observed. The potential advantages of transcatheter perimembranous ventricular septal defect closure over surgery warrant further exploration of differing transcatheter techniques.

Methods: Between August, 2004 and November, 2009, 21 patients had a perimembranous ventricular septal defect closed with transcatheter techniques. Of these, 14 were closed with a muscular amplatz ventricular septal defect device. The median age and weight at device placement were 8 years, ranging from 2 to 19 years, and 18.6 kilograms, ranging from 10 to 21 kilograms, respectively.

Results: There were 25 procedures performed on 23 patients using 21 amplatz ventricular septal defect devices. Median defect size on angiography was 7.8 millimetres, ranging from 4 to 14.3 millimetres, with a median device size of 8 millimetres, ranging from 4 to 18 millimetres, and a defect/device ratio of 1.1, with a range from 0.85 to 1.33. Median procedure time was 100 minutes, with a range from 38 to 235 minutes. Adverse events included device embolisation following haemolysis in one, and new aortic incompetence in another, but there were no cases of heart block. Median follow-up was 41.7 months, with a range from 2 to 71 months.

Conclusions: Evaluating transcatheter closure of perimembranous ventricular septal defect using amplatz ventricular septal defect devices remains important, if a technically feasible method with low and acceptable complication rates is to be identified. Incidence of heart block may be minimised by avoiding oversized devices, using muscular devices, and accepting defeat if an appropriately selected device pulls through. Given the current transcatheter technologies, the closure of perimembranous ventricular septal defects should generally be performed in children when they weigh at least 10 kilograms.

Download full-text PDF

Source
http://dx.doi.org/10.1017/S1047951110002039DOI Listing

Publication Analysis

Top Keywords

ventricular septal
48
septal defect
36
perimembranous ventricular
28
amplatz ventricular
16
closure perimembranous
12
ventricular
12
septal
12
incidence heart
12
defect devices
12
septal defects
12

Similar Publications

Background: In adolescents and adults with tetralogy of Fallot (TOF), right ventricle (RV) electromechanical dyssynchrony (EMD) due to right bundle branch block (RBBB) is associated with reduced exercise capacity and RV dysfunction. While the development of RBBB following surgical repair of tetralogy of Fallot (rTOF) is a frequent sequela, it is not known whether EMD is present in every patient immediately following rTOF. The specific timing of the onset of RBBB following rTOF therefore provides an opportunity to assess whether acute RBBB is associated with the simultaneous acquisition of EMD.

View Article and Find Full Text PDF

Background: This article focuses on the effect of body mass index (BMI) on cardiac structure and function in cases with non-valvular atrial fibrillation (NVAF). Only a few articles have investigated the relationship between BMI and the incidence of left atrial thrombus (LAT) or spontaneous echo contrast (SEC) in cases with NVAF.

Methods: This single-center retrospective study was conducted at The First People's Hospital of Changzhou.

View Article and Find Full Text PDF

Interventricular septal dissection secondary to acute inferior myocardial infarction: case series and literature review.

J Cardiothorac Surg

January 2025

Echocardiography and Vascular Ultrasound Center, The First Affiliated Hospital, Zhejiang University School of Medicine, 79 Qingchun Road, Hangzhou, 310003, China.

Background: Interventricular septal dissection is a critical disease characterized by the separation of the intraventricular septum into two layers, forming an intermediate layer with a cystic cavity that communicates with the root of the aorta or ventricle. It has low morbidity and high mortality rates.

Case Presentation: Case 1: A 58-year-old male with a history of hypertension and smoking presented to a local hospital due to chest tightness and pain for 4 days.

View Article and Find Full Text PDF

Postinfarction ventricular septal rupture (PIVSR) is a rare but serious complication of acute myocardial infarction. Determining how to conduct surgical repair safely is critical. We compared the outcomes of Impella and intra-aortic balloon pump (IABP) implantation during perioperative mechanical circulatory support management in patients with PIVSR (n = 22).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!