Testing whether the barrier of skin samples has sufficient integrity for meaningful measurements of in-vitro chemical permeability is usually required when data are generated for regulatory purposes. Recently, skin integrity has been assessed using LCR databridge measurements, which are reported as resistances determined in either series (SER) or parallel (PAR) modes at a single frequency, typically 100 or 1000Hz. Measurements made at different combinations of mode and frequency are known to differ, although the skin literature reveals confusion over the meaning of these differences and the impact on the interpretation of integrity test results. Here, the theoretical meanings of resistance and capacitance measurements in PAR and SER mode are described and confirmed experimentally. SER-mode resistances are equal to the real part of the complex impedance; whereas, PAR-mode resistances are the inverse of the real part of the admittance. Capacitance measurements reported in SER and PAR modes are similar manipulations of the imaginary parts of the complex impedance and admittance. A large body of data from human cadaver skin is used to show that the PAR-mode resistance and SER-mode capacitance measured at 100Hz are sensitive to skin resistivity, which is the electrical measurement most closely related to skin integrity.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9397369 | PMC |
http://dx.doi.org/10.1016/j.tiv.2011.01.013 | DOI Listing |
Toxicol In Vitro
December 2011
Chemical and Biological Engineering Department, Colorado School of Mines, Golden, CO 80401, USA.
The objective of this study was to quantitatively compare measurements of tritiated water permeability with impedance determined at either 100 or 1000 Hz using an LCR databridge on the same pieces of skin. A previously published expression based on a simple circuit of a parallel resistor and constant phase element (CPE) was used to relate (RPARA) measured at different frequencies to the DC resistance (RskinA) and the steady-state skin permeability of tritiated water (kp). Using this analysis, kp and (RPARA) data from three laboratories were shown to be consistent with each other, and kp and (RskinA) estimated from (RPARA) were linearly correlated.
View Article and Find Full Text PDFToxicol In Vitro
June 2011
Chemical Engineering Department, Colorado School of Mines, Golden, CO 80401, USA.
Testing whether the barrier of skin samples has sufficient integrity for meaningful measurements of in-vitro chemical permeability is usually required when data are generated for regulatory purposes. Recently, skin integrity has been assessed using LCR databridge measurements, which are reported as resistances determined in either series (SER) or parallel (PAR) modes at a single frequency, typically 100 or 1000Hz. Measurements made at different combinations of mode and frequency are known to differ, although the skin literature reveals confusion over the meaning of these differences and the impact on the interpretation of integrity test results.
View Article and Find Full Text PDFToxicol In Vitro
October 2004
The DuPont Company, Haskell Laboratory for Health and Environmental Sciences, P.O. Box 50, Elkton Road, Newark, DE 19714-0050, USA.
Electrical impedance is used to confirm skin integrity for in vitro dermal regulatory testing and as a tool to evaluate skin condition to determine the irritation and corrosion potential of various chemicals and personal care products. In this experiment, samples of dermatomed human skin were mounted onto static diffusion cells (0.64 cm2) maintained at 32 degrees C.