Failure of proximal femoral locking compression plate: a case series.

J Orthop Trauma

Department of Orthopaedic Surgery, NYU Hospital for Joint Diseases, New York, NY, USA.

Published: February 2011

Objectives: The treatment of fractures of the proximal femur has evolved significantly over recent years. The most recent advance is the proximal femoral locking compression plate (PF-LCP; Synthes, West Chester, PA). We present seven failures of the PF-LCP in an attempt to elicit reasons for the failures so that these pitfalls can be avoided in the future.

Design: Retrospective chart review.

Setting: Level I trauma centers, tertiary referral hospitals.

Patients/participants: Ten patients at five institutions.

Intervention: Open reduction and internal fixation with a PF-LCP.

Main Outcome Measurements: Failure mode, time until failure, need for further surgery, bony union, or conversion to arthroplasty.

Results: Of the seven cases, two were acute peritrochanteric fractures, one was a periprosthetic fracture at the site of a prior hip fusion, one was an early failure of a compression hip screw, and three were nonunions. The failure mode was implant fracture in four cases and loss of fixation in three cases resulting from varus collapse and implant cutout. Five of seven failures were within the first 3 weeks (average, 12.4 days). The average time to failure for all cases was 37.9 days (range, 5-175 days). The average patient age was 56.7 years (range, 36-72 years).

Conclusion: Biomechanical studies have shown the PF-LCP to be stronger or equivalent to other fixation methods for fractures of the femoral neck and subtrochanteric femur fractures. The seven failures in our cases may be partially the result of patient factors as well as technical factors; however, there appears to be a high rate of failure even when surgery is performed by experienced and fellowship-trained traumatologists.

Download full-text PDF

Source
http://dx.doi.org/10.1097/BOT.0b013e3181e31cccDOI Listing

Publication Analysis

Top Keywords

proximal femoral
8
femoral locking
8
locking compression
8
compression plate
8
failure mode
8
time failure
8
failure surgery
8
days average
8
failure
7
cases
5

Similar Publications

Background: Currently, there is no clear standard for the surgical options for Evans-Jensen Type IV intertrochanteric femur fractures in elderly patients over 65 years old. This study aims to retrospectively analyze and compare the early postoperative limb function and quality of life of patients treated with total hip arthroplasty (THA) and proximal femoral nail antirotation (PFNA) for this type of fracture. We hypothesize that there is no significant difference in complications and postoperative recovery between the two surgical methods.

View Article and Find Full Text PDF

Purpose: Malalignment of the lower extremity can affect one, two or all three anatomic planes. We hypothesized an influence between the malalignment of the coronal and axial planes.

Methods: A total of 356 lower extremities of 226 patients were included.

View Article and Find Full Text PDF

Purpose: Tranexamic acid is widely accepted for hip fractures but there is no agreement about dose or application method and the use is still off label for hip fractures. The aim of our study was to find the best application method of tranexamic acid in patients with femoral neck fractures comparing total blood loss, hemoglobin and transfusion rate.

Methods: A retrospective single centre cohort study (level I trauma centre) with 2008 patients treated operatively for a proximal femur fracture between January 2016 and January 2022 was performed.

View Article and Find Full Text PDF

This study aimed to summarize the clinical characteristics of acute epiphyseal osteomyelitis and acute metaphyseal osteomyelitis of long bones in children. Data of 43 children with acute osteomyelitis of long bones diagnosed and treated from November 2017 to January 2021 were retrospectively analyzed. Medical records, laboratory results, and MRI were reviewed.

View Article and Find Full Text PDF

Mid-term Clinical and Radiographic Outcomes of the Actis Total Hip System: A Retrospective Study.

Cureus

January 2025

Department of Orthopaedic Surgery, Faculty of Medicine, Dentistry and Pharmaceutical Sciences, Okayama University, Okayama, JPN.

Introduction Implant technology for total hip arthroplasty (THA) was developed to improve hip function and patient satisfaction. Actis (DePuy Synthes, Warsaw, IN, USA) is a short fit-and-fill titanium stem, with a medial-collared and triple-taper (MCTT) geometry, that is fully coated with hydroxyapatite (HA). We evaluated the radiographic and clinical outcomes of the Actis Total Hip System during a mean follow-up of five years.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!