Objective: The purpose was to evaluate the image noise, contrast-to-noise ratio, and radiation dose at indirect multidetector-row computed tomographic venography (CTV) using a low-kilovolt (peak) (kV[p]) high-tube current technique.

Materials And Methods: Thirty patients underwent indirect CTV of the lower extremity at 120 kV(p) with 170 mA and at 80 kV(p) with 426 mA (CT dose index volumes 11.8 and 9.0 mGy, respectively) on a 64-detector CT scanner. We assessed CT numbers, image noise, and contrast-to-noise ratio of deep (femoral and popliteal) veins.

Results: The mean (SD) CT numbers of the femoral and popliteal veins were significantly greater at 80 kV(p) than at 120 kV(p) (142.8 [24.5] and 147.0 [19.4] Hounsfield units [HU], respectively, vs 93.1 [15.5] and 105.0 [14.9] HU, respectively; P < 0.01). The mean (SD) image noise was significantly higher at 80 kV(p) than at 120 kV(p) (11.6 [2.6] and 7.4 [1.4] HU, respectively, vs 8.9 [1.6] and 6.2 [1.5] HU, respectively; P < 0.01). Contrast-to-noise ratios at the femoral and popliteal veins were significantly higher at 80 kV(p) than at 120 kV(p) (6.0 [2.2] and 9.7 [2.4], respectively vs 3.3 [1.8] and 6.9 [2.5], respectively; P < 0.01).

Conclusion: At indirect CTV, 80 kV(p) and a high tube current setting yielded significantly improved image quality at a reduced radiation dose compared with 120 kV(p).

Download full-text PDF

Source
http://dx.doi.org/10.1097/RCT.0b013e3181f5a62eDOI Listing

Publication Analysis

Top Keywords

120 kvp
20
radiation dose
12
image noise
12
femoral popliteal
12
kvp 120
12
kvp
10
low-kilovolt peak
8
high-tube current
8
dose indirect
8
indirect multidetector-row
8

Similar Publications

This study aims to explore the feasibility of applying the "Three-Low" technique (low injection rate, low iodine contrast volume, low radiation dose) in coronary CT angiography (CCTA). We prospectively collected data from 90 patients who underwent CCTA at our hospital between 2021 and 2024. The patients were randomly assigned to either the experimental group (n = 45) or the control group (n = 45).

View Article and Find Full Text PDF

Background: Triple rule-out computed tomography angiography (CTA) provides imaging of the coronary arteries, pulmonary arteries, and thoracic aorta filled with contrast material (CM) to exclude or diagnose the pathologies of these three systems. Although dual rule-out adapted to exclude aortic and pulmonary pathologies. Iodinated CM may result in contrast-induced nephropathy, which lengthens hospital stay.

View Article and Find Full Text PDF

Objective: Biodegradable cardiovascular stents made of thin, low atomic number metals (e.g. Zn, Mg, Fe) are now approved for clinical use.

View Article and Find Full Text PDF

Background: This study aimed to determine whether spectral imaging with dual-energy computed tomography (CT) can improve diagnostic performance for coronary plaque characterization.

Methods And Results: We conducted a retrospective analysis of 30 patients with coronary plaques, using coronary CT angiography (dual-layer CT) and intravascular ultrasound (IVUS) studies. Based on IVUS findings, patients were diagnosed with either vulnerable or stable plaques.

View Article and Find Full Text PDF

Purpose: To investigate the behavior of artificial intelligence (AI) CT-based body composition biomarkers at different virtual monoenergetic imaging (VMI) levels using dual-energy CT (DECT).

Methods: This retrospective study included 88 contrast-enhanced abdominopelvic CTs acquired with rapid-kVp switching DECT. Images were reconstructed into five VMI levels (40, 55, 70, 85, 100 keV).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!