Comparison of anterior segment measurements with rotating Scheimpflug photography and partial coherence reflectometry.

J Cataract Refract Surg

Key Laboratory of Vision Science, School of Optometry and Ophthalmology and Eye Hospital, Wenzhou Medical College, Ministry of Health PR China.

Published: February 2011

AI Article Synopsis

Article Abstract

Purpose: To compare central corneal thickness (CCT), anterior chamber depth (ACD), and keratometry (K) readings measured using optical low-coherence reflectometry (OLCR) biometry and high-resolution rotating Scheimpflug photography.

Setting: Eye Hospital of Wenzhou Medical College, Wenzhou, China.

Design: Comparative case series.

Methods: The CCT, ACD endothelium to lens, ACD epithelium to lens, and K (mean; in flattest meridian; in steepest meridian) were measured 5 times using the LenStar/Biograph OLCR biometer and 3 times with the Pentacam Scheimpflug system in eyes of healthy volunteers. Concordance was evaluated using paired t tests, the Pearson correlation, and Bland-Altman analyses.

Results: The CCT, ACD endothelium to lens, and ACD epithelium to lens measured with the Scheimpflug system were slightly, albeit significantly, higher than with the OLCR biometer (P<.05); the respective 95% limits of agreement (LoA) were -8.2 μm to 15.7 μm, -0.11 to 0.15 mm, and -0.13 to 0.17 mm. However, the Scheimpflug system gave significantly flatter readings for K in the flattest meridian (95% LoA, -0.54 to 0.32 diopters [D]), K in the steepest meridian (95% LoA, -0.63 to 0.45 D), and mean K (95% LoA, -0.53 to 0.33 D) (P<.001). The CCT, ACD, and K readings were all highly correlated between the 2 devices (r >0.95, P<.001).

Conclusions: The CCT and ACD measurements with the OLCR biometer and Scheimpflug system can be used interchangeably in healthy young subjects. However, for K measurements, these devices have wide LoA so may not be interchangeable under certain clinical circumstances.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcrs.2010.08.044DOI Listing

Publication Analysis

Top Keywords

rotating scheimpflug
8
cct acd
8
acd endothelium
8
endothelium lens
8
lens acd
8
acd epithelium
8
epithelium lens
8
olcr biometer
8
scheimpflug system
8
acd
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!