Objective: We wanted to evaluate the efficacy of stent placement using the coaxial technique with a stiff, long introducer sheath in patients with technical failure using an angiographic catheter for the obstructions proximal to the descending colon.

Materials And Methods: Self-expandable metallic stent placement was attempted under fluoroscopy-guidance in 77 consecutive patients who had malignant colorectal obstruction. Stent placement was performed using an angiographic catheter and a guide wire. If the angiographic catheter could not be advanced over the guide wire into the obstructive lesions proximal to the descending colon, then a 6-Fr introducer sheath was used. The technical success rate, the clinical success rate and the complications were analyzed.

Results: Successful stent placement was achieved in 75 of 77 patients (97%). The angiographic catheter failed to advance into the obstructive lesions of 11 patients (M:F = 7:4; mean age, 65.5 years) whose lesions were at the level of the splenic flexure or transverse colon. Therefore, the coaxial technique was implemented in all these 11 patients using a 6-Fr stiff introducer sheath and then the stent placement was successful. There were no complications related to the use of a stiff introducer sheath. Clinical success, which was defined as relief of clinical obstructive bowel symptoms, was obtained within 24 hours in all of patients.

Conclusion: The coaxial technique using a stiff introducer sheath can increase the technical success of fluoroscopy-guided, self-expandable metallic stent placement in patients with colonic obstruction proximal to the descending colon.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3017874PMC
http://dx.doi.org/10.3348/kjr.2011.12.1.107DOI Listing

Publication Analysis

Top Keywords

stent placement
28
introducer sheath
24
coaxial technique
16
proximal descending
16
angiographic catheter
16
descending colon
12
stiff introducer
12
6-fr introducer
8
sheath stent
8
obstructions proximal
8

Similar Publications

Background: Endoscopic ultrasound-guided transmural drainage (EUS-TD) is widely performed to treat postoperative peripancreatic fluid collection (POPFC). Recent reports on EUS-TD lack a consensus on stent selection. This study aimed to assess the efficacy of EUS-TD for POPFC using an external drainage-based approach.

View Article and Find Full Text PDF

Aneurysm dome and vessel pressure measurements with coiling, stent assisted coiling and flow diversion.

Acta Neurochir (Wien)

January 2025

Division of Neurosurgery, Beth Israel Deaconess Medical Center, Harvard Medical School, 110 Francis Street , Boston, MA, 02215, USA.

Background: Variability in long-term endovascular treatment outcomes for intracranial aneurysms has prompted questions regarding the effects of these treatments on aneurysm hemodynamics. Endovascular techniques disrupt aneurysmal blood flow and shear, but their influence on intra-aneurysmal pressure remains unclear. A better understanding of aneurysm pressure effects may aid in predicting outcomes and guiding treatment decisions.

View Article and Find Full Text PDF

Background And Aims: Efficacy and safety of EUS-guided placement of lumen-apposing metal stents (LAMS) has been reported yet advantage of using 15-mm LAMS over 10-mm LAMS yet to be explored.

Methods: International, retrospective, case-matched study of patients with intra-abdominal abscess who underwent EUS-guided drainage with 15-mm (case) and 10-mm (control) LAMS between 03/2019 and 09/2022.

Results: 51 patients underwent EUS-guided drainage using LAMS [15-mm 29 (57%), 10-mm 22 (43%)].

View Article and Find Full Text PDF

Biliary stent occlusion is due, in part, to biofilm formation by bacteria. However, previous culture-based approaches may not have revealed all microorganisms on the surface. Twenty-seven patients underwent endoscopic retrograde biliary drainage for the removal or replacement of plastic biliary stents.

View Article and Find Full Text PDF

Background: Hepatic artery infusion pump (HAIP) chemotherapy is a locoregional treatment for intrahepatic malignancies. HAIPs are surgically implanted, and the catheter tip is typically inserted into a ligated gastroduodenal artery stump. Potential complications at the catheter insertion site include dehiscence, pseudoaneurysm or extravasation, and adjacent hepatic arterial stenosis and thrombosis.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!