In some cases of Intensity-modulated radiotherapy (IMRT) point dose measurement, there exists significant deviation between calculated and measured dose at isocenter, sometimes greater than ±3%. This may be because IMRT fields generate complex profiles at the reference point. The deviation arises due to lack of lateral electronic equilibrium for small fields, and other factors such as leakage and scatter contribution. Measurements were done using 0.125-cc ion chamber and Universal IMRT phantom (both from PTW-Freiburg). The aim is to find a suitable point of measurement for the chamber to avoid discrepancy between calculated and measured dose. Various beam profiles were generated in the plane of the chamber for each field by implementing patient plan on the IMRT phantom. The profiles show that for the fields which are showing deviation, the ion chamber lies in the steep-gradient region. To rectify the problem, the TPS (Treatment Planning System) calculated dose is found out at various points in the measurement plane of the chamber at isocenter. The necessary displacement to the chamber, as noted from the TPS, was given to obtain the optimum result. Twenty cases were studied for optimization, whose percentage deviation was more than ±3%. The results were well within tolerance criteria of ±3% after optimization. The mean percentage deviation value for the 20 cases studied, with standard deviation of 2.33 under 95% confidence interval, was found out to be 2.10% ± 1.14. Those cases that have significant variation even after optimization are further studied with film dosimetry.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3014100PMC
http://dx.doi.org/10.4103/0971-6203.37480DOI Listing

Publication Analysis

Top Keywords

point dose
8
dose measurement
8
intensity-modulated radiotherapy
8
calculated measured
8
measured dose
8
ion chamber
8
imrt phantom
8
plane chamber
8
cases studied
8
optimization percentage
8

Similar Publications

Introduction: Therapeutic drug monitoring (TDM) in inflammatory rheumatic diseases (RMDs) is gaining interest. However, there are unresolved questions about the best practices for implementing TDM effectively in clinical settings.

Objective: The primary objective of this study was to evaluate whether early TDM of adalimumab predicts drug survival at 52 weeks in patients with RMDs.

View Article and Find Full Text PDF

Purpose: BMS-986299 is a first-in-class, NOD-, LRR-, and pyrin-domain containing-3 (NLRP3) inflammasome agonist enhancing adaptive immune and T-cell memory responses.

Materials And Methods: This was a phase-I (NCT03444753) study that assessed the safety and tolerability of intra-tumoral BMS-986299 monotherapy (part 1A) and in combination (part 1B) with nivolumab, and ipilimumab in advanced solid tumors. Reported here are single-center results.

View Article and Find Full Text PDF

Feasibility of very low iodine dose aortoiliac CT angiography using dual-source photon-counting detector CT.

Eur J Radiol

January 2025

Department of Diagnostic and Interventional Radiology, Medical Center-University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, USA. Electronic address:

Purpose: To evaluate the feasibility of aortoiliac CT-Angiography (CTA) using dual-source photon-counting detector (PCD)-CT with minimal iodine dose.

Methods: This IRB-approved, single-center prospective study enrolled patients with indications for aortoiliac CTA from December 2022 to March 2023. All scans were performed using a first-generation dual-source PCD-CT.

View Article and Find Full Text PDF

Objective: The purpose of this study was to present a newly designed 3D-printed personalized model (3D PPM) of a radiofrequency needle guide with a maxillary fixation for gasserian ganglion (GG) puncture.

Methods: Implementation of 3D CT-guided radiofrequency therapy of the GG with and without use of 3D PPM was analyzed. The following parameters were assessed: radiation time, dose area product, air kerma reference point, pain severity during the puncture needle insertion, prosopalgia regression degree (according to visual analog scale) and the severity of facial numbness (according to the Barrow Neurological Institute scale) in the early postoperative period, and postpuncture complications.

View Article and Find Full Text PDF

Introduction: Esketamine has unique advantages in combination with dexmedetomidine for sedation in young children, owing to its sympathetic activity and mild respiratory depression. However, the optimal dose is yet to be determined. In this study, we compared the different doses of intranasal esketamine combined with dexmedetomidine for sedation during transthoracic echocardiography in toddlers.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!